Preoperative proximal tibial bone density, bone microarchitecture, and bone turnover are not associated with postoperative tibial component migration in cemented and cementless medial unicompartmental knee replacements: secondary analyses from a randomized controlled trial
DOI:
https://doi.org/10.2340/17453674.2024.39917Keywords:
Arthroplasty, Histomorphometry, Knee, Micro-CT, RSAAbstract
Background and purpose: Cementless arthroplasty fixation relies on early bone ingrowth and may be poor in patients with low proximal tibial bone density or abnormal bone turnover. We aimed first to describe the baseline bone properties in patients undergoing medial unicompartmental knee replacement (UKR), and second to investigate its association with cemented and cementless tibial component migration until 2 years.
Methods: A subset investigation of 2 patient groups from a 3-armed randomized controlled trial was conducted. There were 26 cemented and 25 cementless medial UKRs with twin-pegged femoral components. Volumetric bone mineral density (vBMD) and microstructure of the excised medial tibial plateau were ascertained with µCT. Bone turnover was estimated using dynamic histomorphometry (eroded surface/bone surface = ES/BS, osteoid surface/bone surface = OS/BS, mineralizing surface/bone surface = MS/BS). Tibial component migration in 4 feature points was followed for 2 years with radiostereometry.
Results: At the 2-year follow-up, the cementless tibial components migrated 0.38 mm (95% confidence interval [CI] 0.14–0.62) total translation more than the cemented components at the posterior feature point. The greatest migration in the cementless group was subsidence at the posterior feature point of 0.66 mm (CI 0.48–0.84) until 6 weeks, and from 3 months the components were stable. Cemented tibial components subsided very little. Between 1- and 2-year follow-ups, no cementless but 4 cemented tibial components revealed continuous migration. OS/BS was half of the ES/BS. No µCT or histomorphometric parameters showed any clinically relevant correlation with tibial component migration at the posterior feature point for either cemented or cementless UKR at 6 weeks’ or 2 years’ follow-up after adjustment for age, BMI, and sex.
Conclusion: Preoperative vBMD, bone turnover, and microstructure were not associated with postoperative tibial component migration of cemented and cementless medial UKR.
Downloads
References
Ben-Shlomo Y, Blom A, Boulton C, Brittain R, Clark E, Dawson-Bowling S, et al. The National Joint Registry 18th Annual Report 2021. London; 2021. PMID: 35072993. Available from https://www.ncbi.nlm.nih.gov/books/NBK576858/
Stratton-Powell A A, Williams S, Tipper J L, Redmond A C, Brockett C L. Mixed material wear particle isolation from periprosthetic tissue surrounding total joint replacements. J Biomed Mater Res B Appl Biomater 2022; 110: 2276-89. doi: 10.1002/jbm.b.35076. DOI: https://doi.org/10.1002/jbm.b.35076
Andersen M R, Winther N S, Lind T, Schroder H M, Flivik G, Petersen M M. Low preoperative BMD Is related to high migration of tibia components in uncemented TKA-92 patients in a combined DEXA and RSA study with 2-year follow-up. J Arthroplasty 2017; 32: 2141-6. doi: 10.1016/j.arth.2017.02.032. DOI: https://doi.org/10.1016/j.arth.2017.02.032
Ryd L, Albrektsson B E, Carlsson L, Dansgard F, Herberts P, Lindstrand A, et al. Roentgen stereophotogrammetric analysis as a predictor of mechanical loosening of knee prostheses. J Bone Joint Surg Br 1995; 77: 377-83. PMID: 7744919 DOI: https://doi.org/10.1302/0301-620X.77B3.7744919
Pijls B G, Valstar E R, Nouta K A, Plevier J W, Fiocco M, Middeldorp S, et al. Early migration of tibial components is associated with late revision: a systematic review and meta-analysis of 21,000 knee arthroplasties. Acta Orthop 2012; 83: 614-624. doi: 10.3109/17453674.2012.747052 DOI: https://doi.org/10.3109/17453674.2012.747052
Koppens D, Rytter S, Dalsgaard J, Sorensen O G, Hansen TB, Stilling M. The effect of bone quality on tibial component migration in medial cemented unicompartmental knee arthroplasty: a prospective cohort study using dual x-ray absorptiometry and radiostereometric analysis. J Arthroplasty 2020; 35: 675-82 e672. doi: 10.1016/j.arth.2019.10.027. DOI: https://doi.org/10.1016/j.arth.2019.10.027
Mosegaard S B, Odgaard A, Madsen F, Romer L, Kristensen P W, Vind T D, et al. Comparison of cementless twin-peg, cemented twin-peg and cemented single-peg femoral component migration after medial unicompartmental knee replacement: a 5-year randomized RSA study. Arch Orthop Trauma Surg 2023; 10.1007/s00402-023-04991-y. doi: 10.1007/s00402-023-04991-y. DOI: https://doi.org/10.1007/s00402-023-04991-y
Horsager K, Madsen F, Odgaard A, Fink Jepsen C, Rømer L, Kristensen P W, et al. Similar polyethylene wear between cemented and cementless Oxford medial UKA: a 5-year follow-up randomized controlled trial on 79 patients using radiostereometry. Acta Orthop 2019; 90: 67-73. doi: 10.1080/17453674.2018.1543757. DOI: https://doi.org/10.1080/17453674.2018.1543757
Dempster D W, Compston J E, Drezner M K, Glorieux F H, Kanis J A, Malluche H, et al. Standardized nomenclature, symbols, and units for bone histomorphometry: a 2012 update of the report of the ASBMR Histomorphometry Nomenclature Committee. J Bone Miner Res 2013; 28: 2-17. doi: 10.1002/jbmr.1805/. DOI: https://doi.org/10.1002/jbmr.1805
Eriksen E F. Normal and pathological remodeling of human trabecular bone: three dimensional reconstruction of the remodeling sequence in normals and in metabolic bone disease. Endocr Rev 1986; 7: 379-408. doi: 10.1210/edrv-7-4-379. DOI: https://doi.org/10.1210/edrv-7-4-379
Sikjaer T, Rejnmark L, Thomsen JS, Tietze A, Bruel A, Andersen G, et al. Changes in 3-dimensional bone structure indices in hypoparathyroid patients treated with PTH(1-84): a randomized controlled study. J Bone Miner Res 2012; 27: 781-8. doi: 10.1002/jbmr.1493 DOI: https://doi.org/10.1002/jbmr.1493
Thomsen J S, Laib A, Koller B, Prohaska S, Mosekilde L, Gowin W. Stereological measures of trabecular bone structure: comparison of 3D micro computed tomography with 2D histological sections in human proximal tibial bone biopsies. J Microsc 2005; 218: 171-179. doi: 10.1111/j.1365-2818.2005.01469.x. DOI: https://doi.org/10.1111/j.1365-2818.2005.01469.x
Kaptein B L, Valstar E R, Stoel B C, Rozing P M, Reiber J H. A new model-based RSA method validated using CAD models and models from reversed engineering. J Biomech 2003; 36: 873-82. doi: 10.1016/s0021-9290(03)00002-2. DOI: https://doi.org/10.1016/S0021-9290(03)00002-2
Valstar E R, Gill R, Ryd L, Flivik G, Borlin N, Kärrholm J. Guidelines for standardization of radiostereometry (RSA) of implants. Acta Orthop 2005; 76: 563-72. doi: 10.1080/17453670510041574. DOI: https://doi.org/10.1080/17453670510041574
Pijls B G, Plevier J W M, Nelissen R. RSA migration of total knee replacements. Acta Orthop 2018; 89: 320-8. doi: 10.1080/17453674.2018.1443635. DOI: https://doi.org/10.1080/17453674.2018.1443635
Kendrick B J, Kaptein B L, Valstar E R, Gill H S, Jackson W F, Dodd C A, et al. Cemented versus cementless Oxford unicompartmental knee arthroplasty using radiostereometric analysis: a randomised controlled trial. Bone Joint J 2015; 97-B: 185-91. doi: 10.1302/0301-620X.97B2.34331. DOI: https://doi.org/10.1302/0301-620X.97B2.34331
Mukaka M M. Statistics corner: a guide to appropriate use of correlation coefficient in medical research. Malawi Med J 2012; 24: 69-71. https://doi.org/10.4236/jwarp.2015.77047. DOI: https://doi.org/10.4236/jwarp.2015.77047
Ryd L. Micromotion in knee arthroplasty: a roentgen stereophotogrammetric analysis of tibial component fixation. Acta Orthop Scand Suppl 1986; 220: 1-80. PMID: 3461667
Qiu S, Divine G, Warner E, Rao S D. Reference intervals for bone histomorphometric measurements based on data from healthy premenopausal women. Calcif Tissue Int 2020; 107: 543-50. doi: 10.1007/s00223-020-00748-6. DOI: https://doi.org/10.1007/s00223-020-00748-6
Klose-Jensen R, Hartlev L B, Boel L W T, Laursen M B, Stengaard-Pedersen K, Keller K K, et al. Subchondral bone turnover, but not bone volume, is increased in early stage osteoarthritic lesions in the human hip joint. Osteoarthritis Cartilage 2015; 23: 2167-73. doi: 10.1016/j.joca.2015.06.001. DOI: https://doi.org/10.1016/j.joca.2015.06.001
Kienapfel H, Sprey C, Wilke A, Griss P. Implant fixation by bone ingrowth. J Arthroplasty 1999; 14: 355-68. doi: 10.1016/s0883-5403(99)90063-3. DOI: https://doi.org/10.1016/S0883-5403(99)90063-3
Mohammad H R, Kennedy J A, Mellon S J, Judge A, Dodd C A, Murray D W. The clinical outcomes of cementless unicompartmental knee replacement in patients with reduced bone mineral density. J Orthop Surg Res 2020; 15: 35. doi: 10.1186/s13018-020-1566-2. DOI: https://doi.org/10.1186/s13018-020-1566-2
Odgaard A, Pedersen C M, Bentzen S M, Jorgensen J, Hvid I. Density changes at the proximal tibia after medial meniscectomy. J Orthop Res 1989; 7: 744-53. doi: 10.1002/jor.1100070517. DOI: https://doi.org/10.1002/jor.1100070517
Li M G, Nilsson K G. Changes in bone mineral density at the proximal tibia after total knee arthroplasty: a 2-year follow-up of 28 knees using dual energy X-ray absorptiometry. J Orthop Res 2000; 18: 40-7. doi: 10.1002/jor.1100180107. DOI: https://doi.org/10.1002/jor.1100180107
Regner L R, Carlsson L V, Karrholm J N, Hansson T H, Herberts P G, Swanpalmer J. Bone mineral and migratory patterns in uncemented total knee arthroplasties: a randomized 5-year follow-up study of 38 knees. Acta Orthop Scand 1999; 70: 603-8. doi: 10.3109/17453679908997850. DOI: https://doi.org/10.3109/17453679908997850
Linde K N, Madsen F, Puhakka K B, Langdahl B L, Soballe K, Krog-Mikkelsen I, et al. preoperative systemic bone quality does not affect tibial component migration in knee arthroplasty: a 2-year radiostereometric analysis of a hundred consecutive patients. J Arthroplasty 2019; 34: 2351-9. doi: 10.3109/17453679908997850. DOI: https://doi.org/10.1016/j.arth.2019.05.019
Li M G, Nilsson K G. The effect of the preoperative bone quality on the fixation of the tibial component in total knee arthroplasty. J Arthroplasty 2000; 15: 744-53. doi: 10.1054/arth.2000.6617. DOI: https://doi.org/10.1054/arth.2000.6617
Petersen M M, Nielsen P T, Lebech A, Toksvig-Larsen S, Lund B. Preoperative bone mineral density of the proximal tibia and migration of the tibial component after uncemented total knee arthroplasty. J Arthroplasty 1999; 14: 77-81. doi: 10.1016/s0883-5403(99)90206-1. DOI: https://doi.org/10.1016/S0883-5403(99)90206-1
Koppens D, Stilling M, Munk S, Dalsgaard J, Rytter S, Sorensen O G, et al. Low implant migration of the SIGMA((R)) medial unicompartmental knee arthroplasty. Knee Surg Sports Traumatol Arthrosc 2018; 26: 1776-85. doi: 10.1007/s00167-017-4782-5. DOI: https://doi.org/10.1007/s00167-017-4782-5
Additional Files
Published
How to Cite
License
Copyright (c) 2024 Mathias Alrø Fichtner Bendtsen, Anders Odgaard, Frank Madsen, Sebastian Breddam Mosegaard, Jesper Skovhus Thomsen, Ellen Margrethe Hauge, Kjeld Søballe, Maiken Stilling
This work is licensed under a Creative Commons Attribution 4.0 International License.