2-year results of an RCT of 2 uncemented isoelastic monoblock acetabular components: lower wear rate with vitamin E blended highly cross-linked polyethylene compared to ultra-high molecular weight polyethylene

Authors

  • Joost H J Van Erp Clinical Orthopedic Research Center—mN, Zeist, the Netherlands; Department of Orthopedic Surgery, Diakonessenhuis, Utrecht, the Netherlands
  • Julie R A Massier Clinical Orthopedic Research Center—mN, Zeist, the Netherlands; Department of Orthopedic Surgery, Diakonessenhuis, Utrecht, the Netherlands
  • Jelle J Halma Clinical Orthopedic Research Center—mN, Zeist, the Netherlands
  • Thom E Snijders Clinical Orthopedic Research Center—mN, Zeist, the Netherlands
  • Arthur De Gast Clinical Orthopedic Research Center—mN, Zeist, the Netherlands; Department of Orthopedic Surgery, Diakonessenhuis, Utrecht, the Netherlands

DOI:

https://doi.org/10.1080/17453674.2020.1730073

Abstract

Background and purpose — The long-term survival of arthroplasty components may be limited by polyethylene wear-related problems such as periprosthetic osteolysis and aseptic loosening. Highly cross-linked polyethylene (HXLPE) blended with vitamin E was introduced to improve oxidative stability and to avoid long-term embrittlement. This study clinically compares the tribological behavior and clinical outcome of vitamin E blended HXLPE with ultrahigh molecular weight polyethylene (UHMWPE) in an isoelastic monoblock cup for uncemented total hip arthroplasty.

Patients and methods — In this randomized controlled trial (RCT), 199 patients were included: 102 patients received the vitamin E blended HXLPE cup, 97 patients the UHMWPE cup. Clinical and radiographic parameters were obtained preoperatively, directly postoperative and at 3, 12, and 24 months. Wear rates were compared using the mean linear femoral head penetration (FHP) rate.

Results — 188 patients (94%) completed the 2-year follow-up. Mean patient satisfaction was higher in the vitamin E blended HXLPE group (8.9 [1]) than in in the control group (8.5 [2], p = 0.03). The Harris Hip Score (HHS) was higher in the vitamin E blended HXLPE group (95 [8]) than in the control group (92 [11], p = 0.3). The FHP rate was lower in the vitamin E blended HXLPE group: 0.046 mm/year compared with 0.056 mm/year in the control group (p = 0.05). No adverse reactions associated with the clinical application of vitamin E blended HXLPE were observed during followup, with an excellent 2-year survival to revision rate of 98% for both cups.

Interpretation — This study shows the superior performance of the HXLPE blended with vitamin E acetabular cup with lower linear femoral head penetration rates and better clinical results compared with the UHMWPE acetabular cup after 2 years.

Downloads

Download data is not yet available.

Downloads

Additional Files

Published

2020-02-26

How to Cite

Erp, J. H. J. V., Massier, J. R. A., Halma, J. J., Snijders, T. E., & Gast, A. D. (2020). 2-year results of an RCT of 2 uncemented isoelastic monoblock acetabular components: lower wear rate with vitamin E blended highly cross-linked polyethylene compared to ultra-high molecular weight polyethylene. Acta Orthopaedica, 91(3), 254–259. https://doi.org/10.1080/17453674.2020.1730073