A randomized study on migration of the Spectron EF and the Charnley flanged 40 cemented femoral stems: secondary analysis of a randomized controlled trial using radiostereometric analysis (RSA) with 10 years’ follow-up
DOI:
https://doi.org/10.2340/17453674.2026.46836Keywords:
Arthroplasty, Hip, Implants, RSAAbstract
Background and purpose: Our primary aim was to compare the 10-year migration of the proximally roughened Spectron EF stem with the Charnley flanged 40 stems in cemented THA. Secondary aims were to assess the influence of polyethylene type on Spectron EF migration, and clinical outcomes.
Methods: In this randomized trial (NCT00698672), 150 patients (70% women; mean age 70 years) underwent cemented THA with either a Charnley stem (n = 30) or a Spectron EF stem (n = 120, 4 groups with different articulating surfaces). The articulations utilized conventional polyethylene (CPE) or highly crosslinked polyethylene (XLPE) cup liners combined with either cobalt-chromium (CoCr) or Oxinium (OxZr) femoral heads. RSA was performed postoperatively and up to 10 years. Migration was analyzed using linear mixed-effects models. Clinical outcome was assessed with the Harris Hip Score (HHS).
Results: After 10 years, 48 patients (14 Charnley, 34 Spectron EF) had evaluable RSA data. Both stems showed posterior translation, subsidence, and retroversion. Retroversion was greater for the Charnley stem than the Spectron EF stem at 3.2° (95% confidence interval [CI] 2.7–3.8) and 1.5° (CI 1.2–1.8), respectively, with a mean difference of 1.7° (CI 1.1– 2.4). The Spectron EF stem had more subsidence at 0.72 mm (CI 0.56–0.88) vs 0.32 mm (CI 0.06–0.57) for the Charnley stem, with a difference of 0.41 mm (CI 0.11–0.71). Spectron EF stems articulating with XLPE had 0.44 mm (CI 0.12–0.76) more subsidence than stems articulating with CPE. HHS improved from 41 (SD 15) to 86 (SD 19) for the Charnley stem and 45 (SD 13) to 86 (SD 15) for the Spectron EF stem with no statistically significant difference between the groups.
Conclusion: After 10 years, the Spectron EF stem showed less retroversion, but more subsidence than the Charnley stem, with otherwise similar migration and clinical outcomes. XLPE articulating stems had more subsidence than CPE with the Spectron EF stem.
Downloads
References
Gruen T A, McNeice G M, Amstutz H C. “Modes of failure” of cemented stem-type femoral components: a radiographic analysis of loosening. Clin Orthop Relat Res 1979(141): 17-27. doi: 10.1097/00003086-197906000-00002. DOI: https://doi.org/10.1097/00003086-197906000-00002
Barrack R L. Early failure of modern cemented stems. J Arthroplasty 2000; 15(8): 1036-50. doi: 10.1054/arth.2000.16498. DOI: https://doi.org/10.1054/arth.2000.16498
Scheerlinck T, Casteleyn P P. The design features of cemented femoral hip implants. J Bone Joint Surg Br 2006; 88(11): 1409-18. doi: 10.1302/0301-620x.88b11.17836. DOI: https://doi.org/10.1302/0301-620X.88B11.17836
Jasty M, Maloney W J, Bragdon C R, O’Connor D O, Haire T, Harris W H. The initiation of failure in cemented femoral components of hip arthroplasties. J Bone Joint Surg Br 1991; 73(4): 551-8. doi: 10.1302/0301-620x.73b4.2071634. DOI: https://doi.org/10.1302/0301-620X.73B4.2071634
Kazi H A, Whitehouse S L, Howell J R, Timperley A J. Not all cemented hips are the same: a register-based (NJR) comparison of taper-slip and composite beam femoral stems. Acta Orthop 2019; 90(3): 214-9. doi: 10.1080/17453674.2019.1582680. DOI: https://doi.org/10.1080/17453674.2019.1582680
Børsheim S, Kristensen T B, Hallan G, Gjertsen J E, Furnes O, Dybvik E, et al. Periprosthetic femoral fracture following hip arthroplasty: which component design and fixation method has the lowest risk of reoperation? Bone Joint J 2025; 107-b(9): 873-84. doi: 10.1302/0301-620x.107b9.Bjj-2024-1706.R1. DOI: https://doi.org/10.1302/0301-620X.107B9.BJJ-2024-1706.R1
Johannessen H G, Hallan G, Kadar T, Fenstad A M, Lygre S H L, Haugan K, et al. Polyethylene wear and cup migration of cemented total hip arthroplasty with femoral heads made of oxidized zirconium, steel, or cobalt chromium: a 10-year secondary analysis from a randomized trial using radiostereometry. Acta Orthop 2024; 95: 578-85. doi: 10.2340/17453674.2024.41945. DOI: https://doi.org/10.2340/17453674.2024.41945
Garellick G, Malchau H, Herberts P. The Charnley versus the Spectron hip prosthesis: clinical evaluation of a randomized, prospective study of 2 different hip implants. J Arthroplasty 1999; 14(4): 407-13. doi: 10.1016/s0883-5403(99)90095-5. DOI: https://doi.org/10.1016/S0883-5403(99)90095-5
Malchau H, Herberts P, Eisler T, Garellick G, Söderman P. The Swedish Total Hip Replacement Register. J Bone Joint Surg Am 2002; 84-A (Suppl 2): 2-20. doi: 10.2106/00004623-200200002-00002. DOI: https://doi.org/10.2106/00004623-200200002-00002
Espehaug B, Furnes O, Engesaeter L B, Havelin L I. 18 years of results with cemented primary hip prostheses in the Norwegian Arthroplasty Register: concerns about some newer implants. Acta Orthop 2009; 80(4): 402-12. doi: 10.3109/17453670903161124. DOI: https://doi.org/10.3109/17453670903161124
Kadar T, Hallan G, Aamodt A, Indrekvam K, Badawy M, Havelin L I, et al. A randomized study on migration of the Spectron EF and the Charnley flanged 40 cemented femoral components using radiostereometric analysis at 2 years. Acta Orthop 2011; 82(5): 538-44. doi: 10.3109/17453674.2011.618914. DOI: https://doi.org/10.3109/17453674.2011.618914
Hopewell S, Chan A W, Collins G S, Hróbjartsson A, Moher D, Schulz K F, et al. CONSORT 2025 statement: updated guideline for reporting randomised trials. BMJ 2025; 389:e081123. doi: 10.1136/bmj-2024-081123. DOI: https://doi.org/10.1136/bmj-2024-081123
Kaptein B L, Pijls B, Koster L, Kärrholm J, Hull M, Niesen A, et al. Guideline for RSA and CT-RSA implant migration measurements: an update of standardizations and recommendations. Acta Orthop 2024; 95: 256-67. doi: 10.2340/17453674.2024.40709. DOI: https://doi.org/10.2340/17453674.2024.40709
Kadar T, Hallan G, Aamodt A, Indrekvam K, Badawy M, Skredderstuen A, et al. Wear and migration of highly cross-linked and conventional cemented polyethylene cups with cobalt chrome or Oxinium femoral heads: a randomized radiostereometric study of 150 patients. J Orthop Res 2011; 29(8): 1222-9. doi: 10.1002/jor.21389. DOI: https://doi.org/10.1002/jor.21389
Harris W H. Traumatic arthritis of the hip after dislocation and acetabular fractures: treatment by mold arthroplasty. An end-result study using a new method of result evaluation. J Bone Joint Surg Am 1969; 51(4): 737-55. doi: 10.2106/00004623-196951040-00012. DOI: https://doi.org/10.2106/00004623-196951040-00012
Christensen R, Ranstam J, Overgaard S, Wagner P. Guidelines for a structured manuscript: statistical methods and reporting in biomedical research journals. Acta Orthop 2023; 94: 243-9. doi: 10.2340/17453674.2023.11656. DOI: https://doi.org/10.2340/17453674.2023.11656
Johanson P E, Antonsson M, Shareghi B, Kärrholm J. Early subsidence predicts failure of a cemented femoral stem with minor design changes. Clin Orthop Relat Res 2016; 474(10): 2221-9. doi: 10.1007/s11999-016-4884-2. DOI: https://doi.org/10.1007/s11999-016-4884-2
Söderlund P, Dahl J, Röhrl S, Nivbrant B, Nilsson K G. 10-year results of a new low-monomer cement: follow-up of a randomized RSA study. Acta Orthop 2012; 83(6): 604-8. doi: 10.3109/17453674.2012.742392. DOI: https://doi.org/10.3109/17453674.2012.742392
Nieuwenhuijse M J, Valstar E R, Kaptein B L, Nelissen R G. The Exeter femoral stem continues to migrate during its first decade after implantation: 10–12 years of follow-up with radiostereometric analysis (RSA). Acta Orthop 2012; 83(2): 129-34. doi: 10.3109/17453674.2012.672093. DOI: https://doi.org/10.3109/17453674.2012.672093
Murray D W, Gulati A, Gill H S. Ten-year RSA-measured migration of the Exeter femoral stem. Bone Joint J 2013; 95-b(5): 605-8. doi: 10.1302/0301-620x.95b5.31330. DOI: https://doi.org/10.1302/0301-620X.95B5.31330
Burston B J, Wood J H, Lewis J R, Prosser G H, Firth L, Wood D J, et al. High failure rates of the Spectron EF stem at a minimum of 10 years’ follow-up. J Arthroplasty 2014; 29(10): 1956-60. doi: 10.1016/j.arth.2014.05.003. DOI: https://doi.org/10.1016/j.arth.2014.05.003
Wolf S, Johannessen A C, Ellison P, Furnes O, Hallan G, Rogg K, et al. Inflammatory tissue reactions around aseptically loose cemented hip prostheses: aa retrieval study of the Spectron EF stem with Reflection All-Poly acetabular cup. J Biomed Mater Res B Appl Biomater 2022; 110(7): 1624-36. doi: 10.1002/jbm.b.35023. DOI: https://doi.org/10.1002/jbm.b.35023
Wroblewski B M, Siney P D, Fleming P A. Wear of the cup in the Charnley LFA in the young patient. J Bone Joint Surg Br 2004; 86(4): 498-503. doi: 10.1302/0301-620x.86b4.14249. DOI: https://doi.org/10.1302/0301-620X.86B4.14249
Additional Files
Published
How to Cite
License
Copyright (c) 2026 Håkon Greve Johannessen, Geir Hallan, Ove Furnes, Anne Marie Fenstad, Stein Håkon Lygre, Kristin Haugan, Thomas Kadar, Mona Badawy, Kari Indrekvam, Arild Aamodt, Paul Johan Høl

This work is licensed under a Creative Commons Attribution 4.0 International License.
PlumX (by Elsevier) is an altmetrics platform that tracks and visualizes the online attention, usage, captures, citations, and social media engagement.