A randomized study on migration of the Spectron EF and the Charnley flanged 40 cemented femoral stems: secondary analysis of a randomized controlled trial using radiostereometric analysis (RSA) with 10 years’ follow-up

Authors

  • Håkon Greve Johannessen Department of Clinical Medicine, Faculty of Medicine, University of Bergen, Bergen, Norway https://orcid.org/0009-0006-0310-1300
  • Geir Hallan Department of Clinical Medicine, Faculty of Medicine, University of Bergen, Bergen; The Norwegian Arthroplasty Register, Department of Orthopaedic Surgery, Haukeland University Hospital, Bergen; Coastal Hospital in Hagavik, Department of Orthopaedic Surgery, Haukeland University Hospital, Bergen, Norway
  • Ove Furnes Department of Clinical Medicine, Faculty of Medicine, University of Bergen, Bergen; The Norwegian Arthroplasty Register, Department of Orthopaedic Surgery, Haukeland University Hospital, Bergen, Norway https://orcid.org/0000-0001-8223-2515
  • Anne Marie Fenstad The Norwegian Arthroplasty Register, Department of Orthopaedic Surgery, Haukeland University Hospital, Bergen, Norway https://orcid.org/0000-0002-6429-0153
  • Stein Håkon Lygre The Norwegian Arthroplasty Register, Department of Orthopaedic Surgery, Haukeland University Hospital, Bergen; Department of Occupational Medicine, Haukeland University Hospital, Bergen
  • Kristin Haugan Department of Orthopaedic Surgery, St. Olavs Hospital, Trondheim, Norway https://orcid.org/0000-0003-3414-4523
  • Thomas Kadar Department of Physical Medicine and Rehabilitation, Haukeland University Hospital, Bergen, Norway
  • Mona Badawy Coastal Hospital in Hagavik, Department of Orthopaedic Surgery, Haukeland University Hospital, Bergen, Norway https://orcid.org/0000-0001-6529-1484
  • Kari Indrekvam Department of Clinical Medicine, Faculty of Medicine, University of Bergen, Bergen; Coastal Hospital in Hagavik, Department of Orthopaedic Surgery, Haukeland University Hospital, Bergen, Norway
  • Arild Aamodt Department of Orthopaedic Surgery, Lovisenberg Diakonal Hospital, Oslo, Norway https://orcid.org/0000-0002-2879-9382
  • Paul Johan Høl Department of Clinical Medicine, Faculty of Medicine, University of Bergen, Bergen; Department of Orthopaedic Surgery, Biomatlab, Haukeland University Hospital, Bergen, Norway https://orcid.org/0000-0002-4216-7891

DOI:

https://doi.org/10.2340/17453674.2026.46836

Keywords:

Arthroplasty, Hip, Implants, RSA

Abstract

Background and purpose: Our primary aim was to compare the 10-year migration of the proximally roughened Spectron EF stem with the Charnley flanged 40 stems in cemented THA. Secondary aims were to assess the influence of polyethylene type on Spectron EF migration, and clinical outcomes.
Methods: In this randomized trial (NCT00698672), 150 patients (70% women; mean age 70 years) underwent cemented THA with either a Charnley stem (n = 30) or a Spectron EF stem (n = 120, 4 groups with different articulating surfaces). The articulations utilized conventional polyethylene (CPE) or highly crosslinked polyethylene (XLPE) cup liners combined with either cobalt-chromium (CoCr) or Oxinium (OxZr) femoral heads. RSA was performed postoperatively and up to 10 years. Migration was analyzed using linear mixed-effects models. Clinical outcome was assessed with the Harris Hip Score (HHS).
Results: After 10 years, 48 patients (14 Charnley, 34 Spectron EF) had evaluable RSA data. Both stems showed posterior translation, subsidence, and retroversion. Retroversion was greater for the Charnley stem than the Spectron EF stem at 3.2° (95% confidence interval [CI] 2.7–3.8) and 1.5° (CI 1.2–1.8), respectively, with a mean difference of 1.7° (CI 1.1– 2.4). The Spectron EF stem had more subsidence at 0.72 mm (CI 0.56–0.88) vs 0.32 mm (CI 0.06–0.57) for the Charnley stem, with a difference of 0.41 mm (CI 0.11–0.71). Spectron EF stems articulating with XLPE had 0.44 mm (CI 0.12–0.76) more subsidence than stems articulating with CPE. HHS improved from 41 (SD 15) to 86 (SD 19) for the Charnley stem and 45 (SD 13) to 86 (SD 15) for the Spectron EF stem with no statistically significant difference between the groups.
Conclusion: After 10 years, the Spectron EF stem showed less retroversion, but more subsidence than the Charnley stem, with otherwise similar migration and clinical outcomes. XLPE articulating stems had more subsidence than CPE with the Spectron EF stem.

Downloads

Download data is not yet available.

References

Gruen T A, McNeice G M, Amstutz H C. “Modes of failure” of cemented stem-type femoral components: a radiographic analysis of loosening. Clin Orthop Relat Res 1979(141): 17-27. doi: 10.1097/00003086-197906000-00002. DOI: https://doi.org/10.1097/00003086-197906000-00002

Barrack R L. Early failure of modern cemented stems. J Arthroplasty 2000; 15(8): 1036-50. doi: 10.1054/arth.2000.16498. DOI: https://doi.org/10.1054/arth.2000.16498

Scheerlinck T, Casteleyn P P. The design features of cemented femoral hip implants. J Bone Joint Surg Br 2006; 88(11): 1409-18. doi: 10.1302/0301-620x.88b11.17836. DOI: https://doi.org/10.1302/0301-620X.88B11.17836

Jasty M, Maloney W J, Bragdon C R, O’Connor D O, Haire T, Harris W H. The initiation of failure in cemented femoral components of hip arthroplasties. J Bone Joint Surg Br 1991; 73(4): 551-8. doi: 10.1302/0301-620x.73b4.2071634. DOI: https://doi.org/10.1302/0301-620X.73B4.2071634

Kazi H A, Whitehouse S L, Howell J R, Timperley A J. Not all cemented hips are the same: a register-based (NJR) comparison of taper-slip and composite beam femoral stems. Acta Orthop 2019; 90(3): 214-9. doi: 10.1080/17453674.2019.1582680. DOI: https://doi.org/10.1080/17453674.2019.1582680

Børsheim S, Kristensen T B, Hallan G, Gjertsen J E, Furnes O, Dybvik E, et al. Periprosthetic femoral fracture following hip arthroplasty: which component design and fixation method has the lowest risk of reoperation? Bone Joint J 2025; 107-b(9): 873-84. doi: 10.1302/0301-620x.107b9.Bjj-2024-1706.R1. DOI: https://doi.org/10.1302/0301-620X.107B9.BJJ-2024-1706.R1

Johannessen H G, Hallan G, Kadar T, Fenstad A M, Lygre S H L, Haugan K, et al. Polyethylene wear and cup migration of cemented total hip arthroplasty with femoral heads made of oxidized zirconium, steel, or cobalt chromium: a 10-year secondary analysis from a randomized trial using radiostereometry. Acta Orthop 2024; 95: 578-85. doi: 10.2340/17453674.2024.41945. DOI: https://doi.org/10.2340/17453674.2024.41945

Garellick G, Malchau H, Herberts P. The Charnley versus the Spectron hip prosthesis: clinical evaluation of a randomized, prospective study of 2 different hip implants. J Arthroplasty 1999; 14(4): 407-13. doi: 10.1016/s0883-5403(99)90095-5. DOI: https://doi.org/10.1016/S0883-5403(99)90095-5

Malchau H, Herberts P, Eisler T, Garellick G, Söderman P. The Swedish Total Hip Replacement Register. J Bone Joint Surg Am 2002; 84-A (Suppl 2): 2-20. doi: 10.2106/00004623-200200002-00002. DOI: https://doi.org/10.2106/00004623-200200002-00002

Espehaug B, Furnes O, Engesaeter L B, Havelin L I. 18 years of results with cemented primary hip prostheses in the Norwegian Arthroplasty Register: concerns about some newer implants. Acta Orthop 2009; 80(4): 402-12. doi: 10.3109/17453670903161124. DOI: https://doi.org/10.3109/17453670903161124

Kadar T, Hallan G, Aamodt A, Indrekvam K, Badawy M, Havelin L I, et al. A randomized study on migration of the Spectron EF and the Charnley flanged 40 cemented femoral components using radiostereometric analysis at 2 years. Acta Orthop 2011; 82(5): 538-44. doi: 10.3109/17453674.2011.618914. DOI: https://doi.org/10.3109/17453674.2011.618914

Hopewell S, Chan A W, Collins G S, Hróbjartsson A, Moher D, Schulz K F, et al. CONSORT 2025 statement: updated guideline for reporting randomised trials. BMJ 2025; 389:e081123. doi: 10.1136/bmj-2024-081123. DOI: https://doi.org/10.1136/bmj-2024-081123

Kaptein B L, Pijls B, Koster L, Kärrholm J, Hull M, Niesen A, et al. Guideline for RSA and CT-RSA implant migration measurements: an update of standardizations and recommendations. Acta Orthop 2024; 95: 256-67. doi: 10.2340/17453674.2024.40709. DOI: https://doi.org/10.2340/17453674.2024.40709

Kadar T, Hallan G, Aamodt A, Indrekvam K, Badawy M, Skredderstuen A, et al. Wear and migration of highly cross-linked and conventional cemented polyethylene cups with cobalt chrome or Oxinium femoral heads: a randomized radiostereometric study of 150 patients. J Orthop Res 2011; 29(8): 1222-9. doi: 10.1002/jor.21389. DOI: https://doi.org/10.1002/jor.21389

Harris W H. Traumatic arthritis of the hip after dislocation and acetabular fractures: treatment by mold arthroplasty. An end-result study using a new method of result evaluation. J Bone Joint Surg Am 1969; 51(4): 737-55. doi: 10.2106/00004623-196951040-00012. DOI: https://doi.org/10.2106/00004623-196951040-00012

Christensen R, Ranstam J, Overgaard S, Wagner P. Guidelines for a structured manuscript: statistical methods and reporting in biomedical research journals. Acta Orthop 2023; 94: 243-9. doi: 10.2340/17453674.2023.11656. DOI: https://doi.org/10.2340/17453674.2023.11656

Johanson P E, Antonsson M, Shareghi B, Kärrholm J. Early subsidence predicts failure of a cemented femoral stem with minor design changes. Clin Orthop Relat Res 2016; 474(10): 2221-9. doi: 10.1007/s11999-016-4884-2. DOI: https://doi.org/10.1007/s11999-016-4884-2

Söderlund P, Dahl J, Röhrl S, Nivbrant B, Nilsson K G. 10-year results of a new low-monomer cement: follow-up of a randomized RSA study. Acta Orthop 2012; 83(6): 604-8. doi: 10.3109/17453674.2012.742392. DOI: https://doi.org/10.3109/17453674.2012.742392

Nieuwenhuijse M J, Valstar E R, Kaptein B L, Nelissen R G. The Exeter femoral stem continues to migrate during its first decade after implantation: 10–12 years of follow-up with radiostereometric analysis (RSA). Acta Orthop 2012; 83(2): 129-34. doi: 10.3109/17453674.2012.672093. DOI: https://doi.org/10.3109/17453674.2012.672093

Murray D W, Gulati A, Gill H S. Ten-year RSA-measured migration of the Exeter femoral stem. Bone Joint J 2013; 95-b(5): 605-8. doi: 10.1302/0301-620x.95b5.31330. DOI: https://doi.org/10.1302/0301-620X.95B5.31330

Burston B J, Wood J H, Lewis J R, Prosser G H, Firth L, Wood D J, et al. High failure rates of the Spectron EF stem at a minimum of 10 years’ follow-up. J Arthroplasty 2014; 29(10): 1956-60. doi: 10.1016/j.arth.2014.05.003. DOI: https://doi.org/10.1016/j.arth.2014.05.003

Wolf S, Johannessen A C, Ellison P, Furnes O, Hallan G, Rogg K, et al. Inflammatory tissue reactions around aseptically loose cemented hip prostheses: aa retrieval study of the Spectron EF stem with Reflection All-Poly acetabular cup. J Biomed Mater Res B Appl Biomater 2022; 110(7): 1624-36. doi: 10.1002/jbm.b.35023. DOI: https://doi.org/10.1002/jbm.b.35023

Wroblewski B M, Siney P D, Fleming P A. Wear of the cup in the Charnley LFA in the young patient. J Bone Joint Surg Br 2004; 86(4): 498-503. doi: 10.1302/0301-620x.86b4.14249. DOI: https://doi.org/10.1302/0301-620X.86B4.14249

Published

2026-11-12

How to Cite

Johannessen, H. G., Hallan, G., Furnes, O., Fenstad, A. M., Lygre, S. H., Haugan, K., … Høl, P. J. (2026). A randomized study on migration of the Spectron EF and the Charnley flanged 40 cemented femoral stems: secondary analysis of a randomized controlled trial using radiostereometric analysis (RSA) with 10 years’ follow-up. Acta Orthopaedica, 97, 814–820. https://doi.org/10.2340/17453674.2026.46836

Issue

Section

Publications

Categories

PlumX (by Elsevier) is an altmetrics platform that tracks and visualizes the online attention, usage, captures, citations, and social media engagement.