Similar 2 year migration patterns for a shortened and a distally reduced stem design in uncemented total hip arthroplasty: a randomized controlled trial using radiostereometric analysis
DOI:
https://doi.org/10.2340/17453674.2026.46551Keywords:
Anterior approach, Arthroplasty, Hip, Implants, Minimal invasive surgeryAbstract
Background and purpose: The primary objective was to compare subsidence and retroversion in the first 2 postoperative years of 2 uncemented hip stem designs in primary minimally invasive anterior-supine surgical (MIS-ASI) total hip arthroplasty: a shortened hip stem (Microplasty) and a standard-length stem with reduced distal width (Reduced Distal). Secondary objectives were to compare migration patterns, patient-reported outcome measures (PROMs), clinical and radiological scores, and the occurrence of (serious) adverse events up to 2 years postoperatively.
Methods: 50 patients were randomized in this non-blinded, single-surgeon, industry-funded, radiostereometric analysis study (RSA) (ClinicalTrials.gov NCT03409666). Stem migration was measured relative to directly postoperative at 6 weeks, 1 year, and 2 years postoperatively, using model-based RSA. Linear mixed-effect modelling was used to compare migration. Primary outcomes were subsidence and retroversion at 2 years postoperatively. Secondary outcomes were migration, PROM questionnaires, clinical and radiological scores, and the presence of adverse events in the first 2 postoperative years.
Results: Migration of 15 Microplasty and 20 Reduced Distal stems showed comparable initial migration between direct and 6 weeks postoperative RSA, stabilizing thereafter. Predominant migration directions were subsidence and retroversion with mean values at 2 years of 1.47 mm (95% confidence interval [CI] 0.32–2.61) and 2.57° (CI 1.64–3.50) for the Microplasty stem and 2.35 mm (CI 1.36–3.34) and 1.59° (CI 0.79–2.39) for the Reduced Distal stem. At 6 weeks postoperatively, posterior tilt about the medial–lateral axis was different between stem designs, with 0.07° (CI –0.27 to 0.40) for the Microplasty and 0.48° (CI 0.19–0.77) for the Reduced Distal designs. Secondary outcomes were not clinically different between the stem designs.
Conclusion: We showed that initial subsidence and retroversion for both stem designs in MIS-ASI surgery was present, although likely not clinically relevant. There was no statistically significant difference in subsidence and retroversion between the stem designs. Although initial migration varied substantially between patients and the Reduced Distal design showed more posterior tilting, both designs maintained a stable position from 6 weeks to 2 years postoperatively.
Downloads
References
Yan L, Ge L, Dong S, Saluja K, Li D, Reddy K S, et al. Evaluation of comparative efficacy and safety of surgical approaches for total hip arthroplasty: a systematic review and network meta-analysis. JAMA Netw Open 2023; 6(1): e2253942. doi: 10.1001/jamanetworkopen.2022.53942.
Van Oldenrijk J, Molleman J, Klaver M, Poolman R W, Haverkamp D. Revision rate after short-stem total hip arthroplasty: a systematic review of 49 studies. Acta Orthop 2014; 85(3): 250-8. doi: 10.3109/17453674.2014.908343.
Van Veghel M H W, Hannink G, Van Oldenrijk J, Van Steenbergen L N, Schreurs B W. A comparison of uncemented short versus standard stem length in total hip arthroplasty: results from the Dutch Arthroplasty Register. Acta Orthop 2023; 94: 330-5. doi: 10.2340/17453674.2023.13652.
Molli R G, Lombardi A V Jr, Berend K R, Adams J B, Sneller M A. A short tapered stem reduces intraoperative complications in primary total hip arthroplasty. Clin Orthop Relat Res 2012; 470(2): 450-61. doi: 10.1007/s11999-011-2068-7.
Lombardi A V Jr, Manocchio A G, Berend K R, Morris M J, Adams J B. Early experience with a short, tapered titanium porous plasma sprayed stem with updated design. Surg Technol Int 2018; 32: 233-8.
Gallart X, Fernandez-Valencia J A, Rios G, Bori G, Riba J, Munoz-Mahamud E, et al. Early clinical and radiological outcomes for the Taperloc Complete Microplasty stem. Eur J Orthop Surg Traumatol 2019; 29(3): 619-24. doi: 10.1007/s00590-018-2341-5.
Kok R Y, Koster L A, Kaptein B L, Fiocco M, Keizer S B. A model-based radiostereometric analysis (RSA) randomized control trial evaluating the stability of the cementless Taperloc hip stem: the TapHip study. Acta Orthop 2022; 93: 212-21. doi: 10.2340/17453674.2021.1127.
Kaptein B L, Pijls B, Koster L, Karrholm J, Hull M, Niesen A, et al. Guideline for RSA and CT-RSA implant migration measurements: an update of standardizations and recommendations. Acta Orthop 2024; 95: 256-67. doi: 10.2340/17453674.2024.40709.
Van der Voort P, Pijls B G, Nieuwenhuijse M J, Jasper J, Fiocco M, Plevier J W M, et al. Early subsidence of shape-closed hip arthroplasty stems is associated with late revision: a systematic review and meta-analysis of 24 RSA studies and 56 survival studies. Acta Orthop 2015; 86(5): 575-85. doi: 10.3109/17453674.2015.1043832.
Van der Water L, Righolt C H, Gascoyne T, Kaptein B L, Nelissen R, Pijls B. Migration patterns of uncemented femoral stems in hip replacement: a systematic review and meta-analysis of clinical radiostereometric analysis cohort studies. Acta Orthop 2026; 97: 247-54. doi: 10.2340/17453674.2026.45412.
Kaptein B L, Valstar E R, Stoel B C, Reiber H C, Nelissen R G. Clinical validation of model-based RSA for a total knee prosthesis. Clin Orthop Rel Res 2007; (464): 205-9. doi: 10.1097/BLO.0b013e318157laa5.
Prins A H, Kaptein B L, Stoel B C, Nelissen R, Reiber J H C, Valstar E R. Handling modular hip implants in model-based RSA: combined stem–head models. J Biomech 2008; 41(14): 2912-7. doi: 10.1016/j.jbiomech.2008.07.035.
Van der Lelij T J N, Koster L A, Marang-van de Mheen P J, Toksvig-Larsen S, Nelissen R, Kaptein B L. Influence of marker-selection method in radiostereometric analysis of total knee arthroplasty on tibial baseplate migration patterns: a secondary analysis of a randomized controlled trial with 5-year follow-up. Acta Orthop 2024; 95: 157-65. doi: 10.2340/17453674.2024.40184.
Valstar E R, Gill R, Ryd L, Flivik G, Borlin N, Karrholm J. Guidelines for standardization of radiostereometry (RSA) of implants. Acta Orthop 2005; 76(4): 563-72. doi: 10.1080/17453670510041574.
Kaptein B L, Valstar E R, Stoel B C, Rozing P M, Reiber J H. A new type of model-based Roentgen stereophotogrammetric analysis for solving the occluded marker problem. J Biomech 2005; 38(11): 2330-4. doi: 10.1016/j.jbiomech.2004.09.018.
Noble P C, Alexander J W, Lindahl L J, Yew D T, Granberry W M, Tullos H S. The anatomic basis of femoral component design. Clin Orthop Relat Res 1988; (235): 148-65.
Ghera S, Pavan L. The DePuy Proxima hip: a short stem for total hip arthroplasty. Early experience and technical considerations. Hip Int 2009; 19(3): 215-20. doi: 10.1177/112070000901900305.
Gruen T A, McNeice G M, Amstutz H C. “Modes of failure” of cemented stem-type femoral components: a radiographic analysis of loosening. Clin Orthop Relat Res 1979; (141): 17-27.
Engh C A, Bobyn J D, Glassman A H. Porous-coated hip replacement: the factors governing bone ingrowth, stress shielding, and clinical results. J Bone Joint Surg Br 1987; 69(1): 45-55. doi: 10.1302/0301-620X.69B1.3818732.
Ranstam J, Turkiewicz A, Boonen S, Van Meirhaeghe J, Bastian L, Wardlaw D. Alternative analyses for handling incomplete follow-up in the intention-to-treat analysis: the randomized controlled trial of balloon kyphoplasty versus non-surgical care for vertebral compression fracture (FREE). BMC Med Res Methodol 2012; 12. doi: Artn 35 10.1186/1471-2288-12-35.
Nieuwenhuijse M J, Vehmeijer S B W, Mathijsen N M C, Keizer S B. Fixation of the short global tissue-sparing hip stem. Bone Joint J 2020; 102-B(6): 699-708. doi: 10.1302/0301-620X.102B6.BJJ-2019-1026.R2.
Klein L J, Puretic G, Mohaddes M, Karrholm J. Similar clinical results and early subsidence between the Collum Femoris Preserving and the Corail stem: a randomized radiostereometric study of 77 hips with 2 years’ follow-up. Acta Orthop 2019; 90(3): 202-8. doi: 10.1080/17453674.2019.1577344.
de Waard S, Sierevelt I N, Jonker R, Hoornenborg D, van der Vis H M, Kerkhoffs G, et al. The migration pattern and initial stability of the Optimys short stem in total hip arthroplasty: a prospective 2-year follow-up study of 33 patients with RSA. Hip Int 2020: 1120700020901844. doi: 10.1177/1120700020901844.
Evans S. When and how can endpoints be changed after initiation of a randomized clinical trial? PLoS Clin Trials 2007; 2(4): e18. doi: 10.1371/journal.pctr.0020018.
Nazari-Farsani S, Vuopio M, Loyttyniemi E, Aro H T. Contributing factors to the initial femoral stem migration in cementless total hip arthroplasty of postmenopausal women. J Biomech 2021; 117: 110262. doi: 10.1016/j.jbiomech.2021.110262.
Aro H T, Engelke K, Mattila K, Loyttyniemi E. Volumetric bone mineral density in cementless total hip arthroplasty in postmenopausal women: effects on primary femoral stem stability and clinical recovery. J Bone Joint Surg Am 2021; 103(12): 1072-82. doi: 10.2106/JBJS.20.01614.
Additional Files
Published
How to Cite
License
Copyright (c) 2026 Lennard Koster, Bart L Kaptein, Esmeralda Goorden, Constant A M P Bell, Rob G H H Nelissen

This work is licensed under a Creative Commons Attribution 4.0 International License.