Association of antioxidant-added highly cross-linked polyethylene on revision risk: a registry-based study of 198,073 total hip replacements from the Australian Orthopaedic Association National Joint Replacement Registry between 2014 and 2023

Authors

  • Peter L Lewis Wakefield Orthopaedic Clinic, University of Adelaide, Adelaide; Australian Orthopaedic Association National Joint Replacement Registry (AOANJRR), Adelaide, Australia https://orcid.org/0000-0003-2641-3754
  • David G Campbell Wakefield Orthopaedic Clinic, University of Adelaide, Adelaide, Australia
  • Peiyao Du South Australian Health and Medical Research Institute (SAHMRI), Adelaide, Australia
  • Helena Oakey South Australian Health and Medical Research Institute (SAHMRI), Adelaide, Australia
  • Richard N de Steiger The University of Melbourne, Epworth Hospital, Melbourne, Victoria, Australia
  • Paul N Smith Australian Orthopaedic Association National Joint Replacement Registry (AOANJRR), Adelaide; Australian National University, Canberra, Australia

DOI:

https://doi.org/10.2340/17453674.2025.45181

Keywords:

Arthroplasty, Hip, Polyethylene, Registry

Abstract

Background and purpose: Adding antioxidant to highly cross-linked polyethylene (XLPE) is proposed to improve oxidation resistance and decrease wear in total hip replacements (THR), but long-term performance is unknown. We aimed to compare the revision rates of THR using cementless acetabular components where the insert was made of either XLPE with antioxidant (AOXLPE) or XLPE, using data from a large national registry.
Methods: The population was THR from the Australian Orthopaedic Association National Joint Replacement Registry (AOANJRR) in the 10-year period 2014–2023 with modular cementless acetabular components and ceramic or metal femoral heads used for osteoarthritis. We compared primary THR using XLPE with antioxidant (AOXLPE) acetabular inserts with XLPE acetabular inserts. The outcome measured was all-cause revision. Cumulative percentage revision (CPR) was calculated using the Kaplan–Meier method, and comparisons made using Cox proportional hazards models.
Results: There were 198,073 THRs, of which 35,309 had AOXLPE inserts. There were 769 and 4,327 revisions with AOXLPE and XLPE inserts, respectively. While there was no early difference, the AOXLPE group had a lower revision rate after 3 years (HR 0.64, 95% confidence interval [CI] 0.48–0.84). When adjusted for multiple factors the AOXLPE group still had a lower revision rate after 3 years (HR 0.63, CI 0.47–0.83). Revisions for loosening, wear-related causes, and fracture were proportionately lower in the AOXLPE group, but no difference was found with revisions for dislocation/instability or infection.
Conclusion: While there was no early difference, THR with AOXLPE acetabular inserts had a lower revision rate after 3 years than XLPE. This suggests a possible clinical benefit using AOXLPE but the difference may, in part, be related to the associated femoral or acetabular components.

Downloads

Download data is not yet available.

References

Learmonth I D, Young C, Rorabeck C. The operation of the century: total hip replacement. Lancet 2007; 370: 1508-19. doi: 10.1016/S0140-6736(07)60457-7. DOI: https://doi.org/10.1016/S0140-6736(07)60457-7

de Steiger R, Lorimer M, Graves S E. Cross-linked polyethylene for total hip arthroplasty markedly reduces revision surgery at 16 years. J Bone Joint Surg Am 2018; 100: 1281-8. doi: 10.2106/jbjs.17.01221. DOI: https://doi.org/10.2106/JBJS.17.01221

Shen C, Tang Z H, Hu J Z, Zou G Y, Xiao R C, Yan D X. Does cross-linked polyethylene decrease the revision rate of total hip arthroplasty compared with conventional polyethylene? A meta-analysis. Orthop Traumatol Surg Res 2014; 100: 745-50. doi: 10.1016/j.otsr.2014.07.015. DOI: https://doi.org/10.1016/j.otsr.2014.07.015

Spece H, Yarbrough R V, Kurtz S M. In vivo performance of vitamin E stabilized polyethylene implants for total hip arthroplasty: a review. J Arthroplasty 2023; 38: 970-9. doi: 10.1016/j.arth.2022.11.010. DOI: https://doi.org/10.1016/j.arth.2022.11.010

Lambert B, Neut D, van der Veen H C, Bulstra S K. Effects of vitamin E incorporation in polyethylene on oxidative degradation, wear rates, immune response, and infections in total joint arthroplasty: a review of the current literature. Int Orthop 2019; 43: 1549-57. doi: 10.1007/s00264-018-4237-8. DOI: https://doi.org/10.1007/s00264-018-4237-8

Massaccesi L, Ragone V, Papini N, Goi G, Corsi Romanelli M M, Galliera E. Effects of vitamin E-stabilized ultra high molecular weight polyethylene on oxidative stress response and osteoimmunological response in human osteoblast. Front Endocrinol (Lausanne) 2019; 10: 203. doi: 10.3389/fendo.2019.00203. DOI: https://doi.org/10.3389/fendo.2019.00203

Kjaergaard K, Ding M, Jensen C, Bragdon C, Malchau H, Andreasen C M, et al. Vitamin E-doped total hip arthroplasty liners show similar head penetration to highly cross-linked polyethylene at five years: a multi-arm randomized controlled trial. Bone Joint J 2020; 102-B: 1303-1310. doi: 10.1302/0301-620X.102B10.BJJ-2020-0138.R1. DOI: https://doi.org/10.1302/0301-620X.102B10.BJJ-2020-0138.R1

Busch A, Jager M, Klebingat S, Baghdadi J, Florkemeier T, Hutter F, et al. Vitamin E-blended highly cross-linked polyethylene liners in total hip arthroplasty: a randomized, multicenter trial using virtual CAD-based wear analysis at 5-year follow-up. Arch Orthop Trauma Surg 2020; 140: 1859-66. doi: 10.1007/s00402-020-03358-x. DOI: https://doi.org/10.1007/s00402-020-03358-x

Collins A K, Sauder N, Nepple C M, Blackburn A Z, Prasad A K, Feder O I, et al. Minimum 7-year follow-up of vitamin E-diffused and highly cross-linked polyethylene liners in total hip arthroplasty: findings from a prospective, international, multicenter study of 977 patients. J Arthroplasty 2023; 38: 2373-8. doi: 10.1016/j.arth.2023.05.007. DOI: https://doi.org/10.1016/j.arth.2023.05.007

Elbardesy H, Yadav H, Rabea M, Guerin S, Harty J. Femoral head penetration in Vitamin-E polyethylene liner versus conventional liners in total hip arthroplasty: systematic review and meta-analysis of randomised control trials. SICOT J 2021; 7: 47. doi: 10.1051/sicotj/2021045. DOI: https://doi.org/10.1051/sicotj/2021045

Li Z, Xiang S, Wu C, Wang Y, Weng X. Vitamin E highly cross-linked polyethylene reduces mid-term wear in primary total hip replacement: a meta-analysis and systematic review of randomized clinical trials using radiostereometric analysis. EFORT Open Rev 2021; 6: 759-70. doi: 10.1302/2058-5241.6.200072. DOI: https://doi.org/10.1302/2058-5241.6.200072

Cheng Q Y, Zhang B F, Wen P F, Wang J, Hao L J, Wang T, et al. Vitamin E-enhanced liners in primary total hip arthroplasty: a systematic review and meta-analysis. Biomed Res Int 2021; 2021: 3236679. doi: 10.1155/2021/3236679. DOI: https://doi.org/10.1155/2021/3236679

Lewis PL, Gill D R, McAuliffe M J, McDougall C, Stoney J D, Vertullo C J, et al. Hip, Knee and Shoulder Annual Report 2024. Australian Orthopaedic Association National Joint Replacement Registry. Available from: https://aoanjrr.sahmri.com/documents/10180/1798900/AOANJRR+2024+Annual+Report.pdf/9d0bfe03-2282-8fc8-a424-b8d9abb82b1f?t=1727666185313. (Accessed 13 April 2025). doi: 10.25310/GLOL7776. DOI: https://doi.org/10.25310/GLOL7776

Bildik C, Kahraman H C, Saygi B. Vitamin E-added highly cross-linked polyethylene decreases the risk of osteolysis in an in vivo arthroplasty model. Cureus 2023; 15: e34955. doi: 10.7759/cureus.34955. DOI: https://doi.org/10.7759/cureus.34955

Massier J R A, Van Erp J H J, Snijders T E, Gast A. A vitamin E blended highly cross-linked polyethylene acetabular cup results in less wear: 6-year results of a randomized controlled trial in 199 patients. Acta Orthop 2020; 91: 705-10. doi: 10.1080/17453674.2020.1807220. DOI: https://doi.org/10.1080/17453674.2020.1807220

Nielson T, Owens G, Miller B, Meneghini E, Deckard E R, Meneghini R M. Large femoral heads in total hip arthroplasty with vitamin E highly cross-linked polyethylene: head penetration rates compared to highly cross-linked polyethylene. J Arthroplasty 2022; 37: S685-S691. doi: 10.1016/j.arth.2022.01.075. DOI: https://doi.org/10.1016/j.arth.2022.01.075

Lamo-Espinosa J M, Mariscal G, Gomez-Alvarez J, San-Julian M. Vitamin E cup cross-linked polyethylene in total hip arthroplasty: a meta-analysis of randomized controlled trials. J Am Acad Orthop Surg 2024; 32: e1153-e1165. doi: 10.5435/JAAOS-D-23-01119. DOI: https://doi.org/10.5435/JAAOS-D-23-01119

Wyatt M C, Roberton A, Foxall-Smi M, Beswick A D, Kunutsor S K, Whitehouse M R. Does vitamin E highly-crosslinked polyethylene convey an advantage in primary total hip replacement? A systematic review and meta-analysis. Hip Int 2020; 30: 598-608. doi: 10.1177/1120700019858335. DOI: https://doi.org/10.1177/1120700019858335

Paxton E W, Inacio M C, Namba R S, Love R, Kurtz S M. Metal-on-conventional polyethylene total hip arthroplasty bearing surfaces have a higher risk of revision than metal-on-highly crosslinked polyethylene: results from a US registry. Clin Orthop Relat Res 2015; 473: 1011-21. doi: 10.1007/s11999-014-4105-9. DOI: https://doi.org/10.1007/s11999-014-4105-9

Johanson P E, Furnes O, Ivar Havelin L, Fenstad A M, Pedersen A B, Overgaard S, et al. Outcome in design-specific comparisons between highly crosslinked and conventional polyethylene in total hip arthroplasty. Acta Orthop 2017; 88: 363-9. doi: 10.1080/17453674.2017.1307676. DOI: https://doi.org/10.1080/17453674.2017.1307676

Hemmila M, Laaksonen I, Matilainen M, Eskelinen A, Haapakoski J, Puhto A P, et al. Implant survival of 2,723 vitamin E-infused highly crosslinked polyethylene liners in total hip arthroplasty: data from the Finnish Arthroplasty Register. Acta Orthop 2021; 92: 316-22. doi: 10.1080/17453674.2021.1879513. DOI: https://doi.org/10.1080/17453674.2021.1879513

Kjaergaard K, Varnum C, Ding M, Overgaard S. Revision risk of total hip arthroplasty with vitamin E doped liners: results from the Danish Hip Arthroplasty Register. J Arthroplasty 2022; 37: 1136-42. doi: 10.1016/j.arth.2022.02.007. DOI: https://doi.org/10.1016/j.arth.2022.02.007

Published

2026-01-23

How to Cite

Lewis, P. L., Campbell, D. G., Du, P., Oakey, H., de Steiger, R. N., & Smith, P. N. (2026). Association of antioxidant-added highly cross-linked polyethylene on revision risk: a registry-based study of 198,073 total hip replacements from the Australian Orthopaedic Association National Joint Replacement Registry between 2014 and 2023. Acta Orthopaedica, 97, 28–34. https://doi.org/10.2340/17453674.2025.45181