Early cup migration and wear as predictors for later aseptic loosening: a secondary evaluation of a randomized controlled RSA trial on cemented hip arthroplasties with 18-year follow-up
DOI:
https://doi.org/10.2340/17453674.2025.44328Keywords:
Arthroplasty, Hip, RSAAbstract
Background and purpose: There is no clear evidence on whether migration or wear is the best predictor for later acetabular cup loosening. We aimed to investigate whether early wear or migration, measured via radiostereometric analysis (RSA), predicts later cup loosening. We also compared long-term aseptic loosening rates between conventional (CPE) and highly crosslinked polyethylene (XLPE) cups.
Methods: Data was drawn from a randomized controlled trial (RCT) (ClinicalTrials.Gov NCT00698672) of 150 patients receiving cemented total hip arthroplasties (THAs), with 10-year RSA follow-up. 5 groups were assessed based on implant combinations (Charnley or Spectron EF stems with CPE or XLPE cups and CoCr or Oxinium heads). Migration and wear up to 2 years were evaluated against 18-year cup survival using receiver operating characteristic (ROC) curves.
Results: 19 cups (17 CPE, 2 XLPE) were loose at final follow-up. The area under the ROC curve (AUC) was 0.56 (95% confidence interval [CI] 0.40–0.73) for early migration and 0.85 (CI 0.77–0.94) for early polyethylene (PE) wear, with a difference of 0.29 (CI 0.09–0.49). Hazard ratio for loosening was 0.88 (CI 0.20–3.89) for early migration > 0.2 mm and 19.4 (CI 2.55–147) for early wear > 0.2 mm. At 18 years, survival free of aseptic loosening was 65% (CI 48–77) for CPE and 96% (CI 85–99) for XLPE cups, with a 9-fold higher risk of loosening for CPE.
Conclusion: Early polyethylene wear, not migration, predicted long-term cup loosening. XLPE showed superior long-term performance over CPE with less wear, cup loosening, and revision.
Downloads
References
Kärrholm J, Herberts P, Hultmark P, Malchau H, Nivbrant B, Thanner J. Radiostereometry of hip prostheses: review of methodology and clinical results. Clin Orthop Relat Res 1997; (344): 94-110. doi: 10.1097/00003086-199711000-00011. DOI: https://doi.org/10.1097/00003086-199711000-00011
Madanat R, Mäkinen T J, Aro H T, Bragdon C, Malchau H. Adherence of hip and knee arthroplasty studies to RSA standardization guidelines: a systematic review. Acta Orthop 2014; 85(5): 447-55. doi: 10.3109/17453674.2014.934187. DOI: https://doi.org/10.3109/17453674.2014.934187
Pijls B G, Nieuwenhuijse M J, Fiocco M, Plevier J W, Middeldorp S, Nelissen R G, et al. Early proximal migration of cups is associated with late revision in THA: a systematic review and meta-analysis of 26 RSA studies and 49 survival studies. Acta Orthop 2012; 83(6): 583-91. doi: 10.3109/17453674.2012.745353. DOI: https://doi.org/10.3109/17453674.2012.745353
Callary S A, Solomon L B, Holubowycz O T, Campbell D G, Munn Z, Howie D W. Wear of highly crosslinked polyethylene acetabular components. Acta Orthop 2015; 86(2): 159-68. doi: 10.3109/17453674.2014.972890. DOI: https://doi.org/10.3109/17453674.2014.972890
Dumbleton J H, Manley M T, Edidin A A. A literature review of the association between wear rate and osteolysis in total hip arthroplasty. J Arthroplasty 2002; 17(5): 649-61. doi: 10.1054/arth.2002.33664. DOI: https://doi.org/10.1054/arth.2002.33664
Wilkinson J M, Hamer A J, Stockley I, Eastell R. Polyethylene wear rate and osteolysis: critical threshold versus continuous dose–response relationship. J Orthop Res 2005; 23(3): 520-5. doi: 10.1016/j.orthres.2004.11.005. DOI: https://doi.org/10.1016/j.orthres.2004.11.005
Malak T T, Broomfield J A, Palmer A J, Hopewell S, Carr A, Brown C, et al. Surrogate markers of long-term outcome in primary total hip arthroplasty: a systematic review. Bone Joint Res 2016; 5(6): 206-14. doi: 10.1302/2046-3758.56.2000568. DOI: https://doi.org/10.1302/2046-3758.56.2000568
Cho C H, Pijls B G, Abrahams J M, Roerink A, Katembwe R, Baker A, et al. Migration patterns of acetabular cups: a systematic review and meta-analysis of RSA studies. Acta Orthop 2023; 94: 626-34. doi: 10.2340/17453674.2023.24580. DOI: https://doi.org/10.2340/17453674.2023.24580
Schulz K F, Altman D G, Moher D. CONSORT 2010 Statement: updated guidelines for reporting parallel group randomised trials. BMC Med 2010; 8: 18. doi: 10.1186/1741-7015-8-18. DOI: https://doi.org/10.1186/1741-7015-8-18
Kadar T, Hallan G, Aamodt A, Indrekvam K, Badawy M, Skredderstuen A, et al. Wear and migration of highly cross-linked and conventional cemented polyethylene cups with cobalt chrome or Oxinium femoral heads: a randomized radiostereometric study of 150 patients. J Orthop Res 2011; 29(8): 1222-9. doi: 10.1002/jor.21389. DOI: https://doi.org/10.1002/jor.21389
Jonsson B A, Kadar T, Havelin L I, Haugan K, Espehaug B, Indrekvam K, et al. Oxinium modular femoral heads do not reduce polyethylene wear in cemented total hip arthroplasty at five years: a randomised trial of 120 hips using radiostereometric analysis. Bone Joint J 2015; 97-B(11): 1463-9. doi: 10.1302/0301-620x.97b11.36137. DOI: https://doi.org/10.1302/0301-620X.97B11.36137
Johannessen H G, Hallan G, Kadar T, Fenstad A M, Lygre S H L, Haugan K, et al. Polyethylene wear and cup migration of cemented total hip arthroplasty with femoral heads made of oxidized zirconium, steel, or cobalt chromium: a 10-year secondary analysis from a randomized trial using radiostereometry. Acta Orthop 2024; 95: 578-85. doi: 10.2340/17453674.2024.41945. DOI: https://doi.org/10.2340/17453674.2024.41945
Hardinge K. The direct lateral approach to the hip. J Bone Joint Surg Br 1982; 64(1): 17-19. doi: 10.1302/0301-620x.64b1.7068713. DOI: https://doi.org/10.1302/0301-620X.64B1.7068713
Valstar E R, Gill R, Ryd L, Flivik G, Börlin N, Kärrholm J. Guidelines for standardization of radiostereometry (RSA) of implants. Acta Orthopaedica 2005; 76(4): 563-72. doi: 10.1080/17453670510041574. DOI: https://doi.org/10.1080/17453670510041574
Lydersen S. ROC-kurver og diagnostiske tester. Tidsskr Nor Laegeforen 2018; 138(15). doi: 10.4045/tidsskr.18.0542. DOI: https://doi.org/10.4045/tidsskr.18.0542
Norwegian Arthroplasty Register Annual Report 2024. Available from: https://www.helse-bergen.no/48d1eb/contentassets/9f19d57711ee4e60815d6b89e8e8472b/report2024.pdf
Stauffer R N. Ten-year follow-up study of total hip replacement. J Bone Joint Surg Am 1982; 64(7): 983-90. doi: 10.2106/00004623-198264070-00003. DOI: https://doi.org/10.2106/00004623-198264070-00003
Hoo Z H, Candlish J, Teare D. What is an ROC curve? Emerg Med J 2017; 34(6): 357-9. doi: 10.1136/emermed-2017-206735. DOI: https://doi.org/10.1136/emermed-2017-206735
DeLong E R, DeLong D M, Clarke-Pearson D L. Comparing the areas under two or more correlated receiver operating characteristic curves: a nonparametric approach. Biometrics 1988; 44(3): 837-45. doi: 10.2307/2531595. DOI: https://doi.org/10.2307/2531595
Hauptfleisch J, Glyn-Jones S, Beard D J, Gill H S, Murray D W. The premature failure of the Charnley Elite-Plus stem: a confirmation of RSA predictions. J Bone Joint Surg Br 2006; 88(2): 179-83. doi: 10.1302/0301-620x.88b2.17055. DOI: https://doi.org/10.1302/0301-620X.88B2.17055
Nieuwenhuijse M J, Valstar E R, Kaptein B L, Nelissen R G. The Exeter femoral stem continues to migrate during its first decade after implantation: 10–12 years of follow-up with radiostereometric analysis (RSA). Acta Orthop 2012; 83(2): 129-34. doi: 10.3109/17453674.2012.672093. DOI: https://doi.org/10.3109/17453674.2012.672093
Kärrholm J, Borssén B, Löwenhielm G, Snorrason F. Does early micromotion of femoral stem prostheses matter? 4–7-year stereoradiographic follow-up of 84 cemented prostheses. J Bone Joint Surg Br 1994; 76(6): 912-17. doi: 10.1302/0301-620X.76B6.7983118. DOI: https://doi.org/10.1302/0301-620X.76B6.7983118
Kaptein B L, Pijls B, Koster L, Kärrholm J, Hull M, Niesen A, et al. Guideline for RSA and CT-RSA implant migration measurements: an update of standardizations and recommendations. Acta Orthop 2024; 95: 256-67. doi: 10.2340/17453674.2024.40709. DOI: https://doi.org/10.2340/17453674.2024.40709
McKellop H, Shen F W, Lu B, Campbell P, Salovey R. Effect of sterilization method and other modifications on the wear resistance of acetabular cups made of ultra-high molecular weight polyethylene: a hip-simulator study. J Bone Joint Surg Am 2000; 82(12): 1708-25. doi: 10.2106/00004623-200012000-00004. DOI: https://doi.org/10.2106/00004623-200012000-00004
Harris W H. Wear and periprosthetic osteolysis: the problem. Clin Orthop Relat Res 2001; (393): 66-70. doi: 10.1097/00003086-200112000-00007. DOI: https://doi.org/10.1097/00003086-200112000-00007
Additional Files
Published
How to Cite
License
Copyright (c) 2025 Håkon Greve Johannessen, Geir Hallan, Thomas Kadar, Stein Atle Lie, Stein Håkon Låstad Lygre, Anne Marie Fenstad, Kristin Haugan, Paul Johan Høl, Mona Badawy, Benedikt Jonsson, Kari Indrekvam, Arild Aamodt, Ove Furnes

This work is licensed under a Creative Commons Attribution 4.0 International License.