Polyethylene wear and cup migration of cemented total hip arthroplasty with femoral heads made of oxidized zirconium, steel, or cobalt chromium: a 10-year secondary analysis from a randomized trial using radiostereometry

Authors

  • Håkon Greve Johannessen Department of Clinical Medicine, Faculty of Medicine, University of Bergen, Bergen, Norway https://orcid.org/0009-0006-0310-1300
  • Geir Hallan Department of Clinical Medicine, Faculty of Medicine, University of Bergen, Bergen; Norwegian Arthroplasty Register, Department of Orthopaedic Surgery, Haukeland University Hospital, Bergen; Coastal Hospital in Hagavik, Department of Orthopaedic Surgery, Haukeland University Hospital, Bergen, Norway
  • Thomas Kadar Department of Physical Medicine and Rehabilitation, Haukeland University Hospital, Bergen, Norway
  • Anne Marie Fenstad Norwegian Arthroplasty Register, Department of Orthopaedic Surgery, Haukeland University Hospital, Bergen, Norway https://orcid.org/0000-0002-6429-0153
  • Stein Håkon Låstad Lygre Norwegian Arthroplasty Register, Department of Orthopaedic Surgery, Haukeland University Hospital, Bergen; Department of Occupational Medicine, Haukeland University Hospital, Bergen, Norway
  • Kristin Haugan Department of Orthopaedic Surgery, St. Olavs Hospital, Trondheim, Norway https://orcid.org/0000-0003-3414-4523
  • Paul Johan Høl Department of Clinical Medicine, Faculty of Medicine, University of Bergen, Bergen; Department of Orthopaedic Surgery, Biomatlab, Haukeland University Hospital, Bergen, Norway https://orcid.org/0000-0002-4216-7891
  • Mona Badawy Coastal Hospital in Hagavik, Department of Orthopaedic Surgery, Haukeland University Hospital, Bergen, Norway
  • Benedikt Jonsson Landshospitali University Hospital, Reykjavik, Iceland
  • Kari Indrekvam Department of Clinical Medicine, Faculty of Medicine, University of Bergen, Bergen; Coastal Hospital in Hagavik, Department of Orthopaedic Surgery, Haukeland University Hospital, Bergen, Norway
  • Arild Aamodt Department of Orthopaedic Surgery, Lovisenberg Diakonal Hospital, Oslo, Norway https://orcid.org/0000-0002-2879-9382
  • Ove Furnes Department of Clinical Medicine, Faculty of Medicine, University of Bergen, Bergen; Norwegian Arthroplasty Register, Department of Orthopaedic Surgery, Haukeland University Hospital, Bergen, Norway https://orcid.org/0000-0001-8223-2515

DOI:

https://doi.org/10.2340/17453674.2024.41945

Keywords:

Cemented Total Hip Arthroplasty, Highly crosslinked polyethylene, Oxinium, Radiostereometric Analysis

Abstract

Background and purpose: We aimed to evaluate polyethylene (PE) wear, cup migration, and clinical outcome over 10 years in total hip arthroplasties (THA) using different articulations.
Methods: This is a secondary analysis of 150 patients randomized into 5 groups, using different articulations: Charnley/Charnley Ogee for steel and conventional polyethylene (CPE), or Spectron EF/Reflection with either CPE or highly cross linked polyethylene (XLPE) cups, paired with heads made of either cobalt-chromium (CoCr) or oxidized zirconium (OxZr). All cups were cemented. Patients underwent repeated radiostereometric analysis (RSA) measurements for up to 10 years to assess wear and migration. Clinical outcome was assessed using Harris Hip Score (HHS).
Results: After 10 years, the XLPE cups demonstrated low wear rates: 0.08 mm (95% confidence interval [CI] –0.11 to 0.26 mm) with CoCr heads and 0.06 mm (CI –0.14 to 0.26 mm) with OxZr heads, with a mean difference of 0.01 mm (CI –0.26 to 0.29 mm). In contrast, CPE cups exhibited significantly more wear: 1.35 mm (CI 1.16 to 1.55 mm) with CoCr heads and 1.68 mm (CI 1.44 to 1.92 mm) with OxZr heads, with a mean difference of 0.33 mm (CI 0.02 to 0.64 mm). The Charnley/Ogee group (CPE) showed PE wear of 0.34 mm (CI 0.12 to 0.56 mm). The CPE groups with OxZr and CoCr heads had 0.67 mm (CI 0.38 to 0.96 mm) and 0.35 mm (CI 0.09 to 0.61 mm) greater proximal migration respectively than the corresponding XLPE groups. HHS was similar across all groups.
Conclusion: We found no significant advantage of OxZr over CoCr heads in reducing wear or migration. XLPE demonstrated a major reduction in wear as well as a reduction in cup migration compared with CPE. Charnley performed better than the other CPE cups in terms of PE wear and cup migration. No differences in clinical outcome were found.

Downloads

Download data is not yet available.

References

Harris W H. Wear and periprosthetic osteolysis: the problem. Clin Orthop Relat Res 2001(393): 66-70. doi: 10.1097/00003086-200112000-00007.

McKellop H, Shen F W, Lu B, Campbell P, Salovey R. Effect of sterilization method and other modifications on the wear resistance of acetabular cups made of ultra-high molecular weight polyethylene: a hip-simulator study. J Bone Joint Surg Am 2000; 82(12): 1708-25. doi: 10.2106/00004623-200012000-00004.

Muratoglu O K, Bragdon C R, O’Connor D O, Jasty M, Harris W H. A novel method of cross-linking ultra-high-molecular-weight polyethylene to improve wear, reduce oxidation, and retain mechanical properties. Recipient of the 1999 HAP Paul Award. J Arthroplasty 2001; 16(2): 149-60. doi: 10.1054/arth.2001.20540.

Dorr L D, Wan Z, Shahrdar C, Sirianni L, Boutary M, Yun A. Clinical performance of a Durasul highly cross-linked polyethylene acetabular liner for total hip arthroplasty at five years. J Bone Joint Surg Am 2005; 87(8): 1816-21. doi: 10.2106/jbjs.D.01915.

Digas G, Kärrholm J, Thanner J, Herberts P. 5-year experience of highly cross-linked polyethylene in cemented and uncemented sockets: two randomized studies using radiostereometric analysis. Acta Orthop 2007; 78(6): 746-54. doi: 10.1080/17453670710014518.

Malahias M A, Atrey A, Gu A, Chytas D, Nikolaou V S, Waddell J P. Is oxidized zirconium femoral head superior to other bearing types in total hip arthroplasty? A systematic review and meta-analysis. J Arthroplasty 2019; 34(8): 1844-52. doi: 10.1016/j.arth.2019.03.072.

Kadar T, Hallan G, Aamodt A, Indrekvam K, Badawy M, Skredderstuen A, et al. Wear and migration of highly cross-linked and conventional cemented polyethylene cups with cobalt chrome or Oxinium femoral heads: a randomized radiostereometric study of 150 patients. J Orthop Res 2011; 29(8): 1222-9. doi: 10.1002/jor.21389.

Cho C H, Pijls B G, Abrahams J M, Roerink A, Katembwe R, Baker A, et al. Migration patterns of acetabular cups: a systematic review and meta-analysis of RSA studies. Acta Orthop 2023; 94: 626-34. doi: 10.2340/17453674.2023.24580.

Valstar E R, Gill R, Ryd L, Flivik G, Börlin N, Kärrholm J. Guidelines for standardization of radiostereometry (RSA) of implants. Acta Orthop 2005; 76(4): 563-72. doi: 10.1080/17453670510041574.

Callary S A, Solomon L B, Holubowycz O T, Campbell D G, Munn Z, Howie D W. Wear of highly crosslinked polyethylene acetabular components. Acta Orthop 2015; 86(2): 159-68. doi: 10.3109/17453674.2014.972890.

Harris W H. Traumatic arthritis of the hip after dislocation and acetabular fractures: treatment by mold arthroplasty. An end-result study using a new method of result evaluation. J Bone Joint Surg Am 1969; 51(4): 737-55. doi: 10.2106/00004623-196951040-00012

Jonsson B A, Kadar T, Havelin L I, Haugan K, Espehaug B, Indrekvam K, et al. Oxinium modular femoral heads do not reduce polyethylene wear in cemented total hip arthroplasty at five years: a randomised trial of 120 hips using radiostereometric analysis. Bone Joint J 2015; 97-b(11): 1463-9. doi: 10.1302/0301-620x.97b11.36137.

Juszczak E, Altman D G, Hopewell S, Schulz K. Reporting of multi-arm parallel-group randomized trials: extension of the CONSORT 2010 Statement. JAMA 2019; 321(16): 1610-20. doi: 10.1001/jama.2019.3087.

Kaptein B L, Pijls B, Koster L, Kärrholm J, Hull M, Niesen A, et al. Guideline for RSA and CT-RSA implant migration measurements: an update of standardizations and recommendations. Acta Orthop 2024; 95: 256-67. doi: 10.2340/17453674.2024.40709.

Dumbleton J H, Manley M T, Edidin A A. A literature review of the association between wear rate and osteolysis in total hip arthroplasty. J Arthroplasty 2002; 17(5): 649-61. doi: 10.1054/arth.2002.33664.

Digas G, Thanner J, Nivbrant B, Röhrl S, Ström H, Kärrholm J. Increase in early polyethylene wear after sterilization with ethylene oxide: radiostereometric analyses of 201 total hips. Acta Orthop Scand 2003; 74(5): 531-41. doi: 10.1080/00016470310017910.

Wolf S, Johannessen A C, Ellison P, Furnes O, Hallan G, Rogg K, et al. Inflammatory tissue reactions around aseptically loose cemented hip prostheses: a retrieval study of the Spectron EF stem with Reflection All-Poly acetabular cup. J Biomed Mater Res B Appl Biomater 2022; 110(7): 1624-36. doi: 10.1002/jbm.b.35023.

Salipas A, Poole A S, Teeter M G, Somerville L E, Naudie D D, McCalden R W. A ten-year radiostereometric analysis of polyethylene wear between oxidized zirconium and cobalt chrome articulations in total hip arthroplasty. J Arthroplasty 2022; 37(7S): S692-S696. doi: 10.1016/j.arth.2022.02.099.

Davis E T, Pagkalos J, Kopjar B. Effect of bearing surface on survival of cementless and hybrid total hip arthroplasty: study of data in the National Joint Registry for England, Wales, Northern Ireland and the Isle of Man. JB JS Open Access 2020; 5(2): e0075. doi: 10.2106/jbjs.Oa.19.00075.

Peters R M, Van Steenbergen L N, Stevens M, Rijk P C, Bulstra S K, Zijlstra W P. The effect of bearing type on the outcome of total hip arthroplasty. Acta Orthop 2018; 89(2): 163-9. doi: 10.1080/17453674.2017.1405669.

Wilkinson J M, Hamer A J, Stockley I, Eastell R. Polyethylene wear rate and osteolysis: critical threshold versus continuous dose-response relationship. J Orthop Res 2005; 23(3): 520-5. doi: 10.1016/j.orthres.2004.11.005.

Bragdon C R, Thanner J, Greene M E, Malchau H, Digas G, Harris W H, Kärrholm J. Standing versus supine radiographs in RSA evaluation of femoral head penetration. Clin Orthop Relat Res 2006; 448: 46-51. doi: 10.1097/01.blo.0000224012.50292.67.

Sterne J A C, Savović J, Page M J, Elbers R G, Blencowe N S, Boutron I, et al. RoB 2: a revised tool for assessing risk of bias in randomised trials. BMJ 2019; 366: l4898. doi: 10.1136/bmj.l4898.

Published

2024-09-30

How to Cite

Johannessen, H. G., Hallan, G., Kadar, T., Fenstad, A. M., Lygre, S. H. L., Haugan, K., … Furnes, O. (2024). Polyethylene wear and cup migration of cemented total hip arthroplasty with femoral heads made of oxidized zirconium, steel, or cobalt chromium: a 10-year secondary analysis from a randomized trial using radiostereometry. Acta Orthopaedica, 95, 578–585. https://doi.org/10.2340/17453674.2024.41945

Issue

Section

Articles

Categories