Comparison between model-based RSA and an AI-based CT-RSA: an accuracy study of 30 patients

Authors

DOI:

https://doi.org/10.2340/17453674.2024.35749

Keywords:

AI, Arthroplasty, Artificial intelligence, CT-based migration analysis, CT-RSA, Hip

Abstract

Background and purpose: Radiostereometry (RSA) is the current gold standard for evaluating early implant migration. CT-based migration analysis is a promising method, with fewer handling requirements compared with RSA and no need for implanted bone-markers. We aimed to evaluate agreement between a new artificial intelligence (AI)-based CT-RSA and model-based RSA (MBRSA) in measuring migration of cup and stem in total hip arthroplasty (THA).
Patients and methods: 30 patients with THA for primary osteoarthritis (OA) were included. RSA examinations were performed on the first postoperative day, and at 2 weeks, 3 months, 1, 2, and 5 years after surgery. A low-dose CT scan was done at 2 weeks and 5 years. The agreement between the migration results obtained from MBRSA and AI-based CT-RSA was assessed using Bland–Altman plots.
Results: Stem migration (y-translation) between 2 weeks and 5 years, for the primary outcome measure, was –0.18 (95% confidence interval [CI] –0.31 to –0.05) mm with MBRSA and –0.36 (CI –0.53 to –0.19) mm with AI-based CT-RSA. Corresponding proximal migration of the cup (y-translation) was 0.06 (CI 0.02–0.09) mm and 0.02 (CI –0.01 to 0.05) mm, respectively. The mean difference for all stem and cup comparisons was within the range of MBRSA precision. The AI-based CT-RSA showed no intra- or interobserver variability.
Conclusion: We found good agreement between the AI-based CT-RSA and MBRSA in measuring postoperative implant migration. AI-based CT-RSA ensures user independence and delivers consistent results.

Downloads

Download data is not yet available.

References

Kärrholm J. Roentgen stereophotogrammetry: review of orthopedic applications. Acta Orthop Scand 1989; 60: 491-503. doi: 10.3109/17453678909149328. DOI: https://doi.org/10.3109/17453678909149328

Onsten I, Berzins A, Shott S, Sumner D R. Accuracy and precision of radiostereometric analysis in the measurement of THR femoral component translations: human and canine in vitro models. J Orthop Res 2001; 19: 1162-7. doi: 10.1016/s0736-0266(01)00039-0. DOI: https://doi.org/10.1016/S0736-0266(01)00039-0

Kärrholm J, Borssén B, Löwenhielm G, Snorrason F. Does early micromotion of femoral stem prostheses matter? 4–7-year stereoradiographic follow-up of 84 cemented prostheses. J Bone Joint Surg Br 1994; 76: 912-17. DOI: https://doi.org/10.1302/0301-620X.76B6.7983118

Malchau H, Bragdon C R, Muratoglu O K. The stepwise introduction of innovation into orthopedic surgery: the next level of dilemmas. J Arthroplasty 2011; 26: 825-31. doi: 10.1016/j.arth.2010.08.007. DOI: https://doi.org/10.1016/j.arth.2010.08.007

Pijls B G, Nieuwenhuijse M J, Fiocco M, Plevier J W, Middeldorp S, Nelissen R G, et al. Early proximal migration of cups is associated with late revision in THA: a systematic review and meta-analysis of 26 RSA studies and 49 survival studies. Acta Orthop 2012; 83: 583-91. doi: 10.3109/17453674.2012.745353. DOI: https://doi.org/10.3109/17453674.2012.745353

Valstar E R, Gill R, Ryd L, Flivik G, Börlin N, Kärrholm J. Guidelines for standardization of radiostereometry (RSA) of implants. Acta Orthop 2005; 76: 563-72. doi: 10.1080/17453670510041574. DOI: https://doi.org/10.1080/17453670510041574

Olivecrona H, Maguire G Q, Jr, Noz M E, Zeleznik M P, Kesteris U, Weidenhielm L. A CT method for following patients with both prosthetic replacement and implanted tantalum beads: preliminary analysis with a pelvic model and in seven patients. J Orthop Surg Res 2016; 11: 27. doi: 10.1186/s13018-016-0360-7. DOI: https://doi.org/10.1186/s13018-016-0360-7

Brodén C, Sandberg O, Sköldenberg O, Stigbrand H, Hänni M, Giles J W, et al. Low-dose CT-based implant motion analysis is a precise tool for early migration measurements of hip cups: a clinical study of 24 patients. Acta Orthop 2020; 91: 260-5. doi: 10.1080/17453674.2020.1725345. DOI: https://doi.org/10.1080/17453674.2020.1725345

Geijer M, Rundgren G, Weber L, Flivik G. Effective dose in low-dose CT compared with radiography for templating of total hip arthroplasty. Acta Radiol 2017; 58: 1276-82. doi: 10.1177/0284185117693462. DOI: https://doi.org/10.1177/0284185117693462

Sandberg O H, Kärrholm J, Olivecrona H, Röhrl SM, Sköldenberg O G, Brodén C. Computed tomography-based radiostereometric analysis in orthopedic research: practical guidelines. Acta Orthop 2023; 94: 373-8. doi: 10.2340/17453674.2023.15337. DOI: https://doi.org/10.2340/17453674.2023.15337

Sandberg O, Tholén S, Carlsson S, Wretenberg P. The anatomical SP-CL stem demonstrates a non-progressing migration pattern in the first year: a low dose CT-based migration study in 20 patients. Acta Orthop 2020; 91: 654-9. doi: 10.1080/17453674.2020.1832294. DOI: https://doi.org/10.1080/17453674.2020.1832294

Brodén C, Sandberg O, Olivecrona H, Emery R, Sköldenberg O. Precision of CT-based micromotion analysis is comparable to radiostereometry for early migration measurements in cemented acetabular cups. Acta Orthop 2021; 92(4): 419-23. doi: 10.1080/17453674.2021.1906082. DOI: https://doi.org/10.1080/17453674.2021.1906082

Angelomenos V, Mohaddes M, Itayem R, Shareghi B. Precision of low-dose CT-based micromotion analysis technique for the assessment of early acetabular cup migration compared with gold standard RSA: a prospective study of 30 patients up to 1 year. Acta Orthop 2022; 93: 459-65. doi: 10.2340/17453674.2022.2528. DOI: https://doi.org/10.2340/17453674.2022.2528

Boettner F, Sculco P, Lipman J, Renner L, Faschingbauer M. A novel method to measure femoral component migration by computed tomography: a cadaver study. Arch Orthop Trauma Surg 2016; 136: 857-63. doi: 10.1007/s00402-016-2442-8. DOI: https://doi.org/10.1007/s00402-016-2442-8

Scheerlinck T, Polfliet M, Deklerck R, Van Gompel G, Buls N, Vandemeulebroucke J. Development and validation of an automated and marker-free CT-based spatial analysis method (CTSA) for assessment of femoral hip implant migration: in vitro accuracy and precision comparable to that of radiostereometric analysis (RSA). Acta Orthop 2016; 87: 139-45. doi: 10.3109/17453674.2015.1123569. DOI: https://doi.org/10.3109/17453674.2015.1123569

Brodén C, Olivecrona H, Maguire GQ, Jr, Noz M E, Zeleznik M P, Sköldenberg O. Accuracy and precision of three-dimensional low dose CT compared to standard RSA in acetabular cups: an experimental study. Biomed Res Int 2016; 2016: 5909741. doi: 10.1155/2016/5909741. DOI: https://doi.org/10.1155/2016/5909741

Zhou S, Zhou F, Sun Y, Chen X, Diao Y, Zhao Y, et al. The application of artificial intelligence in spine surgery. Front Surg 2022; 9: 885599. doi: 10.3389/fsurg.2022.885599. DOI: https://doi.org/10.3389/fsurg.2022.885599

Kiernan S, Geijer M, Sundberg M, Flivik G. Effect of symmetrical restoration for the migration of uncemented total hip arthroplasty: a randomized RSA study with 75 patients and 5-year follow-up. J Orthop Surg Res 2020; 15: 225. doi: 10.1186/s13018-020-01736-0. DOI: https://doi.org/10.1186/s13018-020-01736-0

Kiernan S, Kaptein B, Flivik C, Sundberg M, Flivik G. Unexpected varus deformity and concomitant metal ion release and MRI findings of modular-neck hip stems: descriptive RSA study in 75 hips with 8 years’ follow-up. Acta Orthop 2021; 92: 67-73. doi: 10.1080/17453674.2020.1853387. DOI: https://doi.org/10.1080/17453674.2020.1853387

Geijer M, Kiernan S, Sundberg M, Flivik G. Pre- and postoperative offset and femoral neck version measurements and validation using 3D computed tomography in total hip arthroplasty. Acta Radiol Open 2020; 9: 2058460120964911. doi: 10.1177/2058460120964911. DOI: https://doi.org/10.1177/2058460120964911

Esbjörnsson A C, Kiernan S, Mattsson L, Flivik G. Geometrical restoration during total hip arthroplasty is related to change in gait pattern: a study based on computed tomography and three-dimensional gait analysis. BMC Musculoskelet Disord 2021; 22: 369. doi: 10.1186/s12891-021-04226-4. DOI: https://doi.org/10.1186/s12891-021-04226-4

Derbyshire B, Prescott R J, Porter M L. Notes on the use and interpretation of radiostereometric analysis. Acta Orthop 2009; 80: 124-30. doi: 10.1080/17453670902807474. DOI: https://doi.org/10.1080/17453670902807474

Huda W, Magill D, He W. CT effective dose per dose length product using ICRP 103 weighting factors. Med Phys 2011; 38: 1261-5. doi: 10.1118/1.3544350. DOI: https://doi.org/10.1118/1.3544350

Bland J M, Altman D G. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1986; 1: 307-10. DOI: https://doi.org/10.1016/S0140-6736(86)90837-8

Valstar E R, de Jong F W, Vrooman H A, Rozing P M, Reiber J H. Model-based Roentgen stereophotogrammetry of orthopaedic implants. J Biomech 2001; 34: 715-22. doi: 10.1016/s0021-9290(01)00028-8. DOI: https://doi.org/10.1016/S0021-9290(01)00028-8

Shareghi B, Johanson P E, Kärrholm J. Clinical evaluation of model-based radiostereometric analysis to measure femoral head penetration and cup migration in four different cup designs. J Orthop Res 2017; 35: 760-7 . doi: 10.1002/jor.23177. DOI: https://doi.org/10.1002/jor.23177

Kärrholm J, Herberts P, Hultmark P, Malchau H, Nivbrant B, Thanner J. Radiostereometry of hip prostheses: review of methodology and clinical results. Clin Orthop Relat Res 1997: 94-110. DOI: https://doi.org/10.1097/00003086-199711000-00011

Klaassen A D, Schäffer E A, Willigenburg N W, Van Beers L, Scholtes V A B, Van der Hulst V P M, et al. Comparison of early migration patterns between a ceramic and polyethylene liner in uncemented Trabecular Titanium cups: a 2-year randomized controlled trial of 52 hips using radiostereometric analysis. Acta Orthop 2022; 93: 451-8. doi: 10.2340/17453674.2022.2267. DOI: https://doi.org/10.2340/17453674.2022.2267

Kaptein B L, Valstar E R, Spoor C W, Stoel B C, Rozing P M. Model-based RSA of a femoral hip stem using surface and geometrical shape models. Clin Orthop Relat Res 2006; 448: 92-7. doi: 10.1097/01.blo.0000224010.04551.14. DOI: https://doi.org/10.1097/01.blo.0000224010.04551.14

Prins A H, Kaptein B L, Stoel B C, Nelissen R G, Reiber J H, Valstar E R. Handling modular hip implants in model-based RSA: combined stem-head models. J Biomech 2008; 41: 2912-17. doi: 10.1016/j.jbiomech.2008.07.035. DOI: https://doi.org/10.1016/j.jbiomech.2008.07.035

Mettler Jr F A, Huda W, Yoshizumi TT, Mahesh M. Effective doses in radiology and diagnostic nuclear medicine: a catalog. Radiology 2008; 248: 254-63 . doi: 10.1148/radiol.2481071451. DOI: https://doi.org/10.1148/radiol.2481071451

Published

2024-01-29

How to Cite

Christensson, A., Nemati, H. M., & Flivik, G. (2024). Comparison between model-based RSA and an AI-based CT-RSA: an accuracy study of 30 patients. Acta Orthopaedica, 95, 39–46. https://doi.org/10.2340/17453674.2024.35749

Issue

Section

Articles

Categories