Fracture of hip prostheses due to inadequate welding
DOI:
https://doi.org/10.3109/17453678709146382Abstract
Twenty-five patients underwent total hip arthroplasty using a copy of a well-known prosthesis during the period 1974-1976. Five of the prostheses fractured through the neck after 5-12 years. There were no prodromal symptoms and no trauma preceded the fractures. Technical analysis revealed that inadequate welds were responsible.Downloads
Download data is not yet available.
Downloads
Published
1987-01-01
How to Cite
Aspenberg, P., Kolmert, L., Persson, L., & önnerfält, R. (1987). Fracture of hip prostheses due to inadequate welding. Acta Orthopaedica, 58(5), 479–482. https://doi.org/10.3109/17453678709146382
Issue
Section
Articles
License
Acta Orthopaedica (Scandinavica) content is available freely online as from volume 1, 1930. The journal owner owns the copyright for all material published until volume 80, 2009. As of June 2009, the journal has however been published fully Open Access, meaning the authors retain copyright to their work. As of June 2009, articles have been published under CC-BY-NC or CC-BY licenses, unless otherwise specified.