Stability of osteosynthesis in trochanteric fractures: Comparison of three fixation devices in cadavers

Authors

  • Sune Larsson
  • Martin Elloy
  • Lars Ingvar Hansson

DOI:

https://doi.org/10.3109/17453678809149387

Abstract

Stable trochanteric fractures were produced in 12 pairs of human cadaver femora, which were fixed by either a 135° Jewett nail plate, a 135° NoLok™ sliding screw/plate, or a 140° Hansson pin/plate. The bone-implant preparations were then subjected to 20,000 load cycles simulating full weight bearing, measuring the elastic and permanent fracture displacement. In each case, elastic displacement occurred during loading. for the Jewett-stabilized fractures, a steady increased elastic displacement, as well as permanent displacement, was seen throughout the test, causing 2/8 failures when the implant penetrated the femoral head. In the NoLok™-and Hansson-stabilized fractures, there were no failures, both giving better overall stability than the Jewett device with sufficient fracture stability throughout the test.

Downloads

Download data is not yet available.

Downloads

Published

1988-01-01

How to Cite

Larsson, S., Elloy, M., & Hansson, L. I. (1988). Stability of osteosynthesis in trochanteric fractures: Comparison of three fixation devices in cadavers. Acta Orthopaedica, 59(4), 386–390. https://doi.org/10.3109/17453678809149387