Implant holding power of the femoral head
DOI:
https://doi.org/10.3109/17453679209154848Abstract
The holding power against axial pull-out forces of eight different screws implanted in the femoral heads of cadavers was tested. The Olmed screw had the greatest resistance to pull-out forces followed by the recently designed Ullevaal screw with long threads, and the von Bahr screw. There were substantial differences in the holding power in the four sectors of the femoral head. The holding power was strongly correlated with the logarithm of the moment of inertia of the screw threads, which is a function of the area and number of the threads (r = 0.94, P = 0.0002).Downloads
Download data is not yet available.
Downloads
Published
1992-01-01
How to Cite
Benterud, J. G., Husby, T., Graadahl, øivind, & Alho, A. (1992). Implant holding power of the femoral head. Acta Orthopaedica, 63(1), 47–49. https://doi.org/10.3109/17453679209154848
Issue
Section
Articles
License
Acta Orthopaedica (Scandinavica) content is available freely online as from volume 1, 1930. The journal owner owns the copyright for all material published until volume 80, 2009. As of June 2009, the journal has however been published fully Open Access, meaning the authors retain copyright to their work. As of June 2009, articles have been published under CC-BY-NC or CC-BY licenses, unless otherwise specified.