Reproducible radiographs of acetabular prostheses

Authors

  • Jan L Seelen
  • Jon D Bruijn
  • Bettina E Hansen
  • Lucas M Kingma
  • Johan L Bloem

DOI:

https://doi.org/10.3109/17453679408995450

Abstract

35 patients with a smooth, threaded acetabular Mecron type prosthesis were examined with the aid of a table top with wiremarkers and a fixed 30-degree wedge to allow for reproducible positioning. Under fluoroscopic control, pelvic and spot films were made. The inter- and intraobserver variability of anteversion and inclination angle measurements of the prostheses had a standard deviation of less than 1 degree. The method can be applied to other acetabular prostheses as well.

Downloads

Download data is not yet available.

Downloads

Published

1994-01-01

How to Cite

Seelen, J. L., Bruijn, J. D., Hansen, B. E., Kingma, L. M., & Bloem, J. L. (1994). Reproducible radiographs of acetabular prostheses. Acta Orthopaedica, 65(3), 258–262. https://doi.org/10.3109/17453679408995450