Reproducible radiographs of acetabular prostheses
DOI:
https://doi.org/10.3109/17453679408995450Abstract
35 patients with a smooth, threaded acetabular Mecron type prosthesis were examined with the aid of a table top with wiremarkers and a fixed 30-degree wedge to allow for reproducible positioning. Under fluoroscopic control, pelvic and spot films were made. The inter- and intraobserver variability of anteversion and inclination angle measurements of the prostheses had a standard deviation of less than 1 degree. The method can be applied to other acetabular prostheses as well.Downloads
Download data is not yet available.
Downloads
Published
1994-01-01
How to Cite
Seelen, J. L., Bruijn, J. D., Hansen, B. E., Kingma, L. M., & Bloem, J. L. (1994). Reproducible radiographs of acetabular prostheses. Acta Orthopaedica, 65(3), 258–262. https://doi.org/10.3109/17453679408995450
Issue
Section
Articles
License
Acta Orthopaedica (Scandinavica) content is available freely online as from volume 1, 1930. The journal owner owns the copyright for all material published until volume 80, 2009. As of June 2009, the journal has however been published fully Open Access, meaning the authors retain copyright to their work. As of June 2009, articles have been published under CC-BY-NC or CC-BY licenses, unless otherwise specified.