Migration of uncemented, long-stem femoral components in revision hip arthroplasty:A 2-8 year clinical follow-up of 45 cases and radiostereometric analysis of 13 cases
DOI:
https://doi.org/10.3109/17453679508995528Abstract
We reviewed after 48 (24-90) months the clinical results in 45 cases of revision hip arthroplasties where an uncemented, long-stem femoral prosthesis (BIAS®, Zimmer) had been used. A subgroup of 13 cases was followed with radiostereometric analysis (RSA) for 2 years. 3/45 cases had been re-revised, another 12 had unsatisfactory pain scores. The median Harris score was 69 (26-99). 12/13 stems migrated; 11 subsided 4.1 (0.4-7.9) mm, and 8 migrated posteriorly 2.9 (1.9-9.6) mm. The poor clinical results and large migrations speak against the use of this prosthesis in revision hip arthroplasty.Downloads
Download data is not yet available.
Downloads
Published
1995-01-01
How to Cite
Önsten, I., Sanzén, L., Carlsson, Åke, & Besjakov, J. (1995). Migration of uncemented, long-stem femoral components in revision hip arthroplasty:A 2-8 year clinical follow-up of 45 cases and radiostereometric analysis of 13 cases. Acta Orthopaedica, 66(3), 220–224. https://doi.org/10.3109/17453679508995528
Issue
Section
Articles
License
Acta Orthopaedica (Scandinavica) content is available freely online as from volume 1, 1930. The journal owner owns the copyright for all material published until volume 80, 2009. As of June 2009, the journal has however been published fully Open Access, meaning the authors retain copyright to their work. As of June 2009, articles have been published under CC-BY-NC or CC-BY licenses, unless otherwise specified.