Boneloc® cemented total hip prostheses: Loosening in 28/43 cases after 3-38 months
DOI:
https://doi.org/10.3109/17453679508995526Abstract
We report our early results with 43 total hip prostheses (25 Charnley and 18 LMT) inserted during 1991 with Boneloc® cement. The indication was primary arthrosis (38) and rheumatoid arthritis (5). After 18 (3-38) months, 28 stems were loose and 18 hips have been revised 1-4 years after primary surgery.Downloads
Download data is not yet available.
Downloads
Published
1995-01-01
How to Cite
Riegels-nielsen, P., Sørensen, L., Andersen, H. M., & Lindequist, S. (1995). Boneloc® cemented total hip prostheses: Loosening in 28/43 cases after 3-38 months. Acta Orthopaedica, 66(3), 215–217. https://doi.org/10.3109/17453679508995526
Issue
Section
Articles
License
Acta Orthopaedica (Scandinavica) content is available freely online as from volume 1, 1930. The journal owner owns the copyright for all material published until volume 80, 2009. As of June 2009, the journal has however been published fully Open Access, meaning the authors retain copyright to their work. As of June 2009, articles have been published under CC-BY-NC or CC-BY licenses, unless otherwise specified.