Early failure with Boneloc® bone cement: 4/8 femoral stems loose within 3 years

Authors

  • Seppo Suominen

DOI:

https://doi.org/10.3109/17453679508994630

Abstract

I performed 8 hybrid total hip replacement using Boneloc® bone cement. After 32 (28-36) months, 2 reoperations were performed for femoral component loosening and in another 2 cases, the femoral components were loose, as shown on radiographs. The cement was fragmented in the proximal femur and easily removed.

Downloads

Download data is not yet available.

Downloads

Published

1995-01-01

How to Cite