Surface structure of retrieved ICLH femoral cups
DOI:
https://doi.org/10.3109/17453679809000923Abstract
We characterized the surface structure of 21 retrieved femoral ICLH cups, as an indirect measure of surface damage, with an optical scanner specially developed for surface roughness measurements. 3 unused cups were measured and served as controls. The mean insertion time was 10 (5–15) years. 5 areas on each cup were measured, 1 on the top and 4 at the edge. The mean value of 3 different, commonly used, surface roughness parameters was calculated for the edge measurements. We found no correlation between the surface topography and clinical patient data such as age, body-weight, head-shaft angle, implantation time and wear of the polyethylene socket.Downloads
Download data is not yet available.
Downloads
Published
1998-01-01
How to Cite
Wennerberg, A., Rôkkum, M., Reigstad, A., & Albrektsson, T. (1998). Surface structure of retrieved ICLH femoral cups. Acta Orthopaedica, 69(3), 243–247. https://doi.org/10.3109/17453679809000923
Issue
Section
Articles
License
Acta Orthopaedica (Scandinavica) content is available freely online as from volume 1, 1930. The journal owner owns the copyright for all material published until volume 80, 2009. As of June 2009, the journal has however been published fully Open Access, meaning the authors retain copyright to their work. As of June 2009, articles have been published under CC-BY-NC or CC-BY licenses, unless otherwise specified.