Abstract

We evaluated 887 hips in 672 patients with uncemented MicroStructured Omnifit acetabular components for liner locking complications. We found 2 types of radiographic signs of liner locking system complications in 7 hips, developing between 2 and 4 years postoperatively. The incidence of liner locking system complications was 0.8% using this modular acetabular component. We recommend that a patient who has received a total hip arthroplasty including a MicroStructured Omnifit acetabular component should be monitored frequently for radiographic signs of liner locking system complications, especially with a polyethylene thickness of less than 8 mm.

Downloads

Download data is not yet available.

Downloads

Published

2000-01-01

How to Cite

Lee, P.-C., Shih, C.-H., Yen, W.-L., Yang, W.-E., Tu, Y.-K., & Tai, C.-L. (2000). Complications of liner locking system in MicroStructured Omnifit acetabular components: A radiographic evaluation of 887 hips followed for 5-10 years. Acta Orthopaedica, 71(1), 31–33. https://doi.org/10.1080/00016470052943856