Broken wings--common with the Hardinger plug
DOI:
https://doi.org/10.1080/00016470216322Abstract
We used the Hardinger occluder plug in 51 consecutive total hip arthroplasties. In 34 hips, 1 or more of the 12 polyethylene wings fractured and could be washed out from the canal. As many as 7 wings were loose in 2 arthroplasties. Although numerous wings were broken, no failure occurred in occlusion. The placement of loose wings in the femoral canal was studied by cementing a Charnley prosthesis in a sawbone prepared with loose wings. The model was cut into slices. We found that the plastic wings could come into contact with both the bone and prosthesis.Downloads
Download data is not yet available.
Downloads
Published
2002-01-01
How to Cite
Svedmark, P., Wretenberg, P., & Stark, A. (2002). Broken wings--common with the Hardinger plug. Acta Orthopaedica, 73(4), 407–408. https://doi.org/10.1080/00016470216322
Issue
Section
Articles
License
Acta Orthopaedica (Scandinavica) content is available freely online as from volume 1, 1930. The journal owner owns the copyright for all material published until volume 80, 2009. As of June 2009, the journal has however been published fully Open Access, meaning the authors retain copyright to their work. As of June 2009, articles have been published under CC-BY-NC or CC-BY licenses, unless otherwise specified.