Broken wings--common with the Hardinger plug

Authors

  • Per Svedmark
  • Per Wretenberg
  • Andre Stark

DOI:

https://doi.org/10.1080/00016470216322

Abstract

We used the Hardinger occluder plug in 51 consecutive total hip arthroplasties. In 34 hips, 1 or more of the 12 polyethylene wings fractured and could be washed out from the canal. As many as 7 wings were loose in 2 arthroplasties. Although numerous wings were broken, no failure occurred in occlusion. The placement of loose wings in the femoral canal was studied by cementing a Charnley prosthesis in a sawbone prepared with loose wings. The model was cut into slices. We found that the plastic wings could come into contact with both the bone and prosthesis.

Downloads

Download data is not yet available.

Downloads

Published

2002-01-01

How to Cite

Svedmark, P., Wretenberg, P., & Stark, A. (2002). Broken wings--common with the Hardinger plug. Acta Orthopaedica, 73(4), 407–408. https://doi.org/10.1080/00016470216322