Factors predisposing to dislocation of the Thompson hemiarthroplasty: 22 dislocations in 338 patients
DOI:
https://doi.org/10.1080/00016470310013644Abstract
In a series of 338 patients, we have retrospectively analyzed technical and anatomical factors, which may predispose to a dislocation of the Thompson hemiprosthesis. 22 patients (7%) had at least 1 dislocation during the 6-month follow-up. The most significant independent factor predisposing to dislocation was the use of a posterior approach (dislocation rate 16%). We examined the radiographs and data on operations in the 22 patients, using 79 random patients without dislocation as controls. Factors correlating with an increase in the incidence of dislocation were the length of the residual femoral neck > 0.5 cm in short patients (< 165 cm), and considerable change in the postoperative offset of the hip. Acetabular measurements showed no correlation to the dislocation. Our findings suggest that the main factors predicting dislocation are technical and not related to anatomical measurements.Downloads
Download data is not yet available.
Downloads
Published
2003-01-01
How to Cite
Pajarinen, J., Savolainen, V., Tulikoura, I., Lindahl, J., & Hirvensalo, E. (2003). Factors predisposing to dislocation of the Thompson hemiarthroplasty: 22 dislocations in 338 patients. Acta Orthopaedica, 74(1), 45–48. https://doi.org/10.1080/00016470310013644
Issue
Section
Articles
License
Acta Orthopaedica (Scandinavica) content is available freely online as from volume 1, 1930. The journal owner owns the copyright for all material published until volume 80, 2009. As of June 2009, the journal has however been published fully Open Access, meaning the authors retain copyright to their work. As of June 2009, articles have been published under CC-BY-NC or CC-BY licenses, unless otherwise specified.