Pelvic tilt makes acetabular cup navigation inaccurate

Authors

  • Burkhard Lembeck
  • Otto Mueller
  • Patrik Reize
  • Nikolaus Wuelker

DOI:

https://doi.org/10.1080/17453670510041501

Abstract

Background Modern navigation techniques allow precise positioning of the acetabular cup relative to the anterior pelvic plane. Variations in pelvic tilt will affect the resulting spatial orientation of the cup. Methods We measured pelvic tilt in 30 volunteers with an inclinometer combined with an ultrasonographic position measurement system. A mathematical algorithm was developed to calculate the resulting cup position measured on standard radiographs, depending on pelvic tilt. Results Average pelvic tilt at rest was −4° in the lying position and −8° in the standing position, and ranged from −27° to +3°. Pelvic reclination of 1° will lead to functional anteversion of the cup of approximately 0.7°. Interpretation Pelvic tilt makes navigation systems referring to the anterior plane inaccurate.