CT-based migration analysis is more precise than radiostereometric analysis for tibial implants: a phantom study on a porcine cadaver
DOI:
https://doi.org/10.2340/17453674.2023.12306Keywords:
Arthroplasty, Cadaver, Computed tomography, CT-RSA, Diagnostic imaging, Knee arthroplastyAbstract
Background and purpose: Radiostereometric analysis (RSA) is the gold standard for migration analysis, but computed tomography analysis methods (CTRSA) have shown comparable results in other joints. We attempted to validate precision for CT compared with RSA for a tibial implant.
Material and methods: RSA and CT were performed on a porcine knee with a tibial implant. Marker-based RSA, model-based RSA (MBRSA), and CT scans from 2 different manufacturers were compared. CT analysis was performed by 2 raters for reliability evaluation.
Results: 21 double examinations for precision measurements for RSA and CT-based Micromotion Analysis (CTMA) were analysed. Mean (95% confidence interval) precision data for maximum total point motion (MTPM) using marker-based RSA was 0.45 (0.19–0.70) and 0.58 (0.20–0.96) using MBRSA (F-statistic 0.44 [95% CI 0.18–1.1], p = 0.07). Precision data for total translation (TT) for CTMA was 0.08 (0.03–0.12) for the GE scanner and 0.11 (0.04–0.19) for the Siemens scanner (F-statistic 0.37 [0.15–0.91], p = 0.03). When comparing the aforementioned precision for both RSA methods with both CTMA analyses, CTMA was more precise (p < 0.001). The same pattern was seen for other translations and migrations. Mean effective radiation doses were 0.005 mSv (RSA) (0.0048–0.0050) and 0.08 mSv (CT) (0.078–0.080) (p < 0.001). Intra- and interrater reliability were 0.79 (0.75–0.82) and 0.77 (0.72–0.82), respectively.
Conclusion: CTMA is more precise than RSA for migration analysis of a tibial implant, has overall good intra- and interrater reliability but higher effective radiation doses in a porcine cadaver.
Downloads
References
Kärrholm J. Roentgen stereophotogrammetry: review of orthopedic applications. Acta Orthop Scand 1989; 60(4): 491-503. doi: 10.3109/17453678909149328. DOI: https://doi.org/10.3109/17453678909149328
Valstar E R, Gill R, Ryd L, Flivik G, Börlin N, Kärrholm J. Guidelines for standardization of radiostereometry (RSA) of implants. Acta Orthop 2005; 76(4): 563-72. doi: 10.1080/17453670510041574. DOI: https://doi.org/10.1080/17453670510041574
Pijls B G, Valstar E R, Nouta K A, Plevier J W, Fiocco M, Middeldorp S, et al. Early migration of tibial components is associated with late revision: a systematic review and meta-analysis of 21,000 knee arthroplasties. Acta Orthop 2012; 83(6): 614-24. doi: 10.3109/17453674.2012.747052. DOI: https://doi.org/10.3109/17453674.2012.747052
Molt M, Ryd L, Toksvig-Larsen S. A randomized RSA study concentrating especially on continuous migration. Acta Orthop 2016; 87(3): 262-7. doi: 10.3109/17453674.2016.1166876. DOI: https://doi.org/10.3109/17453674.2016.1166876
Implants for Surgery—Roentgen stereophotogrammetric analysis for the assessment of migration of orthopaedic implants. ISO 160872013; 2013.
Hansen L, De Raedt S, Jørgensen P B, Mygind-Klavsen B, Kaptein B, Stilling M. Marker free model-based radiostereometric analysis for evaluation of hip joint kinematics: a validation study. Bone Joint Res 2018; 7(6): 379-87. doi: 10.1302/2046-3758.76.Bjr-2017-0268.R1. DOI: https://doi.org/10.1302/2046-3758.76.BJR-2017-0268.R1
Scheerlinck T, Polfliet M, Deklerck R, Van Gompel G, Buls N, Vandemeulebroucke J. Development and validation of an automated and marker-free CT-based spatial analysis method (CTSA) for assessment of femoral hip implant migration: in vitro accuracy and precision comparable to that of radiostereometric analysis (RSA). Acta Orthop 2016; 87(2): 139-45. doi: 10.3109/17453674.2015.1123569. DOI: https://doi.org/10.3109/17453674.2015.1123569
Brodén C, Giles J W, Popat R, Fetherston S, Olivecrona H, Sandberg O, et al. Accuracy and precision of a CT method for assessing migration in shoulder arthroplasty: an experimental study. Acta Radiol 2020; 61(6): 776-82. doi: 10.1177/0284185119882659. DOI: https://doi.org/10.1177/0284185119882659
Brodén C, Sandberg O, Olivecrona H, Emery R, Sköldenberg O. Precision of CT-based micromotion analysis is comparable to radiostereometry for early migration measurements in cemented acetabular cups. Acta Orthop 2021; 92(4): 419-23. doi: 10.1080/17453674.2021.1906082. DOI: https://doi.org/10.1080/17453674.2021.1906082
Fox A M, Kedgley A E, Lalone E A, Johnson J A, Athwal G S, Jenkyn T R. The effect of decreasing computed tomography dosage on radiostereometric analysis (RSA) accuracy at the glenohumeral joint. J Biomech 2011 ; 44(16): 2847-50. doi: 10.1016/j.jbiomech.2011.08.009. DOI: https://doi.org/10.1016/j.jbiomech.2011.08.009
Boettner F, Sculco P K, Lipman J, Saboeiro G, Renner L, Faschingbauer M. The effect of a low radiation CT protocol on accuracy of CT guided implant migration measurement: a cadaver study. J Orthop Res 2016; 34(4): 725-8. doi: 10.1002/jor.23060. DOI: https://doi.org/10.1002/jor.23060
Brodén C, Sandberg O, Sköldenberg O, Stigbrand H, Hänni M, Giles J W, et al. Low-dose CT-based implant motion analysis is a precise tool for early migration measurements of hip cups: a clinical study of 24 patients. Acta Orthop 2020; 91(3): 260-5. doi: 10.1080/17453674.2020.1725345. DOI: https://doi.org/10.1080/17453674.2020.1725345
Sandgren B, Skorpil M, Nowik P, Olivecrona H, Crafoord J, Weidenhielm L, et al. Assessment of wear and periacetabular osteolysis using dual energy computed tomography on a pig cadaver to identify the lowest acceptable radiation dose. Bone Joint Res 2016; 5(7): 307-13. doi: 10.1302/2046-3758.57.2000566. DOI: https://doi.org/10.1302/2046-3758.57.2000566
Otten V, Maguire Jr G Q, Noz M E, Zeleznik M P, Nilsson K G, Olivecrona H. Are CT scans a satisfactory substitute for the follow-up of RSA migration studies of uncemented cups? A comparison of RSA double examinations and CT datasets of 46 total hip arthroplasties. Biomed Res Int 2017; 2017:3681458. doi: 10.1155/2017/3681458. DOI: https://doi.org/10.1155/2017/3681458
STAndards for the Reporting of Diagnostic accuracy studies (STARD) checklist (cited: 2023 April 1) [database on the Internet]. Equator network. 2015. Available from: https://www.equator-network.org/wp-content/uploads/2015/03/STARD-2015-checklist.pdf.
Tapiovaara M, Siiskonen T. Pcxmc—a pc-based monte carlo program for calculating patient doses in medical x-ray examinations. J ICRU 2005; 5(2): 30. doi: 10.1093/jicru/ndi034. DOI: https://doi.org/10.1093/jicru_ndi034
Saltybaeva N, Jafari M E, Hupfer M, Kalender W A. Estimates of effective dose for CT scans of the lower extremities. Radiology 2014; 273(1): 153-9. doi: 10.1148/radiol.14132903. DOI: https://doi.org/10.1148/radiol.14132903
Niesen A E, Garverick A L, Hull M L. Maximum total point motion of five points versus all points in assessing tibial baseplate stability. J Biomech Eng 2021; 143(11). doi: 10.1115/1.4051347. DOI: https://doi.org/10.1115/1.4051347
Gudnason A, Adalberth G, Nilsson K G, Hailer N P. Tibial component rotation around the transverse axis measured by radiostereometry predicts aseptic loosening better than maximal total point motion. Acta Orthop 2017; 88(3): 282-7. doi: 10.1080/17453674.2017.1297001. DOI: https://doi.org/10.1080/17453674.2017.1297001
Koo T K, Li M Y. A guideline of selecting and reporting intraclass correlation coefficients for reliability research. J Chiropr Med 2016; 15(2): 155-63. doi: 10.1016/j.jcm.2016.02.012. DOI: https://doi.org/10.1016/j.jcm.2016.02.012
Sandberg O, Tholén S, Carlsson S, Wretenberg P. The anatomical SP-CL stem demonstrates a non-progressing migration pattern in the first year: a low dose CT-based migration study in 20 patients. Acta Orthop 2020; 91(6): 654-9. doi: 10.1080/17453674.2020.1832294. DOI: https://doi.org/10.1080/17453674.2020.1832294
Pijls B G, Plevier J W M, Nelissen R G. RSA migration of total knee replacements. Acta Orthop 2018; 89(3): 320-8. doi: 10.1080/17453674.2018.1443635. DOI: https://doi.org/10.1080/17453674.2018.1443635
Angelomenos V, Mohaddes M, Itayem R, Shareghi B. Precision of low-dose CT-based micromotion analysis technique for the assessment of early acetabular cup migration compared with gold standard RSA: a prospective study of 30 patients up to 1 year. Acta Orthop 2022; 93: 459-65. doi: 10.2340/17453674.2022.2528. DOI: https://doi.org/10.2340/17453674.2022.2528
Teeuwisse W, Berting R, Geleijns J. Stralenbelasting bij orthopedische radiologie. Gamma 1998; 4(8-9): 197-200.
No authors listed. Direktoratet for strålevern og atomsikkerhet. 2022 (cited: 2022 Sept 29). Available from: https://radnett.dsa.no/?doc=faq.
Eriksson T, Maguire Jr G Q, Noz M E, Zeleznik M P, Olivecrona H, Shalabi A, et al. Are low-dose CT scans a satisfactory substitute for stereoradiographs for migration studies? A preclinical test of low-dose CT scanning protocols and their application in a pilot patient. Acta Radiol 2019; 60(12): 1643-52. doi: 10.1177/0284185119844166. DOI: https://doi.org/10.1177/0284185119844166
Bojan A J, Bragdon C, Jönsson A, Ekholm C, Kärrholm J. Three-dimensional bone-implant movements in trochanteric hip fractures: precision and accuracy of radiostereometric analysis in a phantom model. J Orthop Res 2015; 33(5): 705-11. doi: 10.1002/jor.22822. DOI: https://doi.org/10.1002/jor.22822
Brodén C, Olivecrona H, Maguire G Q Jr, Noz M E, Zeleznik M P, Sköldenberg O. Accuracy and precision of three-dimensional low dose CT compared to standard RSA in acetabular cups: an experimental study. Biomed Res Int 2016; 2016: 5909741. doi: 10.1155/2016/5909741. DOI: https://doi.org/10.1155/2016/5909741
Fraser A N, Tsukanaka M, Fjalestad T, Madsen J E, Röhrl S M. Model-based RSA is suitable for clinical trials on the glenoid component of reverse total shoulder arthroplasty. J Orthop Res 2018; 36(12): 3299-307. doi: 10.1002/jor.24111. DOI: https://doi.org/10.1002/jor.24111
Kibsgård T J, Røise O, Stuge B, Röhrl S M. Precision and accuracy measurement of radiostereometric analysis applied to movement of the sacroiliac joint. Clin Orthop Relat Res 2012; 470(11): 3187-94. doi: 10.1007/s11999-012-2413-5. DOI: https://doi.org/10.1007/s11999-012-2413-5
Additional Files
Published
How to Cite
License
Copyright (c) 2023 Lars H W Engseth, Anselm Schulz, Are H Pripp, Stephan M H Röhrl, Frank-David Øhrn
This work is licensed under a Creative Commons Attribution-NonCommercial 4.0 International License.