Peter P SCHMITZ 1, Gerjon HANNINK 2, Joey REIJMER 1, Matthijs P SOMFORD 1, and Job L C VAN SUSANTE 1
1 Department of Orthopedics, Rijnstate Hospital, Arnhem; 2 Department of Operating Rooms, Radboud University Medical Center Nijmegen, Nijmegen, The Netherlands
Background and purpose — Trochanteric fractures are often treated using intramedullary fixation. In our institution, the TFN-Advanced Proximal Femoral Nailing System (TFNA) was introduced as replacement for the Gamma Trochanteric Nail (GTN3) for the treatment of these fractures as a result of a hospital-driven change of trauma implant supplier. We compared trochanteric fracture fixation failure rate between these 2 intramedullary nails.
Patients and methods — All trochanteric fractures treated surgically from 2011 to 2019 were retrospectively reviewed for fixation failure. From 2016 only the TFNA was used. Fixation failure was defined as implant cut-out, implant breakage, non-union, malpositioning of the screw/ blade requiring reoperation, new fracture around the nail, or miscellaneous. Propensity score matching was used to balance distribution of covariates and to compare failure rates between TFNA and GTN3 groups. Learning curve analyses were performed.
Results — After exclusion, 797 GTN3s (779 patients) and 542 (536 patients) TFNAs were available for analysis. A higher risk of fixation failure was found in the TFNA group (14%) compared with the GTN3 group (7.0%) (hazard ratio [HR] 2.0, 95% confidence interval [CI] 1.2–3.5). This was mainly attributed to a higher risk of cut-out (HR 2.2; CI 0.9– 5.7), malpositioning (HR 4.7; CI 0.7–34), and new fracture around the nail (HR 4.0; CI 1.0–16). Learning curve analyses indicated no clear learning curve effect.
Interpretation — Failure of fixation increased after a switch from the GTN3 to the TFNA proximal femoral nail for the treatment of trochanteric fractures. Cut-out and malpositioning of the calcar screw or blade appeared to be the most dominant failure mechanisms. Modifications in implant design may have played a role in this increased risk of failure of fixation. In our institution a new implant device was introduced without solid clinical evidence behind it. This study may help to underline the need for medical doctors with a critical and scientific background to be involved in implant choices.
Citation: Acta Orthopaedica 2022; 93: 234–240. DOI http://dx.doi.org/10.2340/17453674.2022.1410.
Copyright: © 2021 The Author(s). This is an Open Access article distributed under the terms of the Creative Commons Attribution-NonCommercial 4.0 International License (https://creativecommons.org/licenses/by-nc/4.0/), allowing third parties to copy and redistribute the material in any medium or format and to remix, transform, and build upon the material for non-commercial purposes, provided proper attribution to the original work.
Submitted: 2021-07-11. Accepted: 2021-12-19. Published: 2022-01-11.
Correspondence: jvansusante@rijnstate.nl
PS and JS designed the study; PS and JR collected the data; GH carried out data analyses; PS, GH, and JS wrote the manuscript; PS, GH, MS, and JS contributed to revision of the manuscript.
Acta thanks Charles Court-Brown and Per Kjærsgaard-Andersen for help with peer review of this study.
Trochanteric fractures are mostly treated with intramedullary fixation. Failures of fixation are not uncommon (4–7%) and lead to reoperation in most cases (1).
The third generation of the Gamma Trochanteric Nail (GTN3; Stryker, Mahwah, NJ, USA) is a widely used implant with established reliability and safety as documented in several clinical studies (2,3). In 2015, the TFN-Advanced Proximal Femoral Nailing System (TFNA) was introduced to the global market, claiming to potentially decrease fixation failure rates through a number of implant changes.
From 2016 the TFNA (DePuy Synthes, West Chester, PA, USA) was used in our institution as replacement for the GTN3 (Stryker) for the treatment of trochanteric fractures. This switch in implant found its origin in a hospital-driven change of trauma implant supplier. We now compare trochanteric fracture fixation failure rates between the 2 intramedullary nails, Gamma Trochanteric Nail (GTN3) and the TFN-Advanced Proximal Femoral Nailing System (TFNA) and evaluate the responsible underlying mechanisms of failure. Since the TFNA was introduced as a new device we also assessed a potential learning curve effect.
Electronic files of all patients treated with an intramedullary nail following a trochanteric fracture between June 2011 and July 2019 were screened. Patients suffering from pathological fractures were excluded. The TFNA was introduced to treat trochanteric fractures in our institution in 2016 whereas the GTN3 had been used since 2004. No GTN3-nail option was available after the introduction of the TFNA in 2016. Patient characteristics regarding age, sex, BMI, smoking status, and ASA score were collected. Length of stay and operation time were noted. Furthermore, type of fracture was defined according to the AO fracture classification system (4). As implant positioning may influence chances of failure of fixation (5), the tip–apex distance (TAD) (6) was assessed on intraoperative anterior–posterior and axial fluoroscopy as proxy measure for implant positioning.
This study assesses radiographic failure of fixation. 6 categories of failure were established and each case was allocated to 1 of the following: 1) implant cut-out, 2) implant breakage, 3) non-union, 4) malpositioning of the screw/blade in the femoral head, 5) new fracture around the nail, and 6) miscellaneous. An implant cut-out was defined as clear migration of hardware in the femoral head and implant breakage as a discontinuity in the device. Non-union was defined as radiographic absence of bone bridging at the fracture site for which reoperation had occurred to remove the distal locking screw. Malpositioning of the screw/blade in the femoral head was defined as an inadequate position requiring early re-intervention according to established consensus from the weekly trauma meeting. A new fracture around the nail was defined as the occurrence of a fracture related to a new fall. Lastly, a small remaining heterogeneous group of failures from collapse of the femoral head, either post-traumatic or from avascular necrosis, was classified as miscellaneous.
The TFNA and GTN3 are roughly similar implants (Figure 1) and are both available in long and short nails. The standard nail has a diameter of 11 mm for both implants and a length of 200 mm in the TFNA versus 180 mm in the GTN3. Both implants have the option for a femoral neck angle of 125 or 130 degrees. For both implants a calcar screw can be used, whereas for the TFNA this screw could now be replaced by a helical blade, which was 1 of the main modifications in implant design (Figure 1). This helical blade was designed to compress bone during insertion to enhance implant anchorage and to reduce the risk of cut-out (7). A blade was used within the TFNA in all patients above 70 years old. For both types of implants we used a similar operative technique. Patients were operated on in a supine position with the use of a traction table and fluoroscopy for closed reduction. All patients received preoperative cefazolin 2,000 mg and nadroparin 2850 IE antiXa till 5 weeks postoperatively. Postoperatively, patients were encouraged to fully weight-bear. With an incidence of around 150 trochanteric fractures per year in our institution all surgeries were performed by or under direct supervision of a certified (orthopedic) trauma surgeon.
Figure 1. A TFN-Advanced Proximal Femoral Nail (TFNA; left) and a third-generation Gamma Trochanteric Nail (GTN3; right). Note the TFNA’s implant modifications: the use of a blade instead of a screw, oblique cut on lateral end of the blade, different bowing, smaller proximal diameter, and different design of the proximal hole.
As a crude analysis, Kaplan–Meier analysis was performed to determine the failure of fixation rates, stratified for implant type (GTN3 or TFNA).
Multiple imputation by chained equations was used to account for missing data. We generated 30 imputed datasets, as current guidance recommends that 1 imputation should be performed per percent of incomplete observations (8). ASA score, BMI, smoking status, and nail size had 2.6%, 5.8%, 30%, and 0.3% missing values, respectively (Table 1).
TFNA | GTN3 | |||
Characteristic | (n = 542) | Missing | (n = 797) | Missing |
Mean age (SD) | 79 (14) | 0 (0) | 80 (12) | 0 (0) |
Female sex | 378 (70) | 0 (0) | 589 (74) | 0 (0) |
ASA score | 0 (0) | 37 (4.6) | ||
1 | 32 (5.9) | N/A | 35 (4.4) | N/A |
2 | 134 (25) | N/A | 315 (40) | N/A |
3 | 276 (51) | N/A | 362 (45) | N/A |
4 | 90 (17) | N/A | 37 (4.6) | N/A |
5 | 10 (1.8) | N/A | 13 (1.6) | N/A |
Mean BMI (SD) | 24.1 (4.4) | 21 (4) | 24.0 (4.2) | 56 (7) |
Smoking, yes | 90 (17) | 0 (0) | 112 (14) | 398 (50) |
Fracture type | 0 (0) | 0 (0) | ||
A1 | 206 (38) | N/A | 290 (36) | N/A |
A2 | 234 (43) | N/A | 326 (41) | N/A |
A3 | 80 (15) | N/A | 131 (16) | N/A |
Subtrochanteric | 22 (4.1) | N/A | 50 (6.3) | N/A |
Screw/blade | 0 (0) | 0 (0 | ||
Screw | 142 (26) | 797 (100) | ||
Blade | 400 (74) | N/A | ||
Size | 0 (0) | 4 (0.5) | ||
Short | 419 (77) | N/A | 620 (78) | N/A |
Long | 123 (23) | N/A | 173 (22) | N/A |
Mean TAD (SD) | 19.8 (6.8) | 28 (5.2) | 17.6 (6.1) | 41 (5.1) |
Days of stay (SD) | 8.8 (6.1) | 0 (0) | 9.6 (7.1) | 0 (0) |
Operation time, minutes (SD) | 45 (22) | 0 (0) | 44 (23) | 0 (0) |
TFNA – TFN-Advanced Proximal Femoral Nailing System GTN3 – Third generation of the Gamma Trochanteric Nail. N/A, not applicable; TAD, tip-apex-distance in mm. |
To adjust for potential confounding baseline characteristics, we matched patients based on their propensity scores. The propensity score was defined as the probability of being treated with either TFNA or GTN3 dependent on a case’s recorded baseline characteristics. Propensity scores were estimated for each imputed dataset, using a logistic regression model with the group (TFNA or GTN3) as the dependent variable in relation to the following baseline characteristics: age, sex, ASA score, BMI, smoking, fracture type, and length of nail (short or long). The selection of which variables to include in our analyses in order to minimize bias was done using directed acyclic graphs based on the approaches described by Shrier and Pearl (9,10). A 1:1 nearest-neighbor matching algorithm was applied without replacement to match cases on their corresponding propensity scores within a caliper of 0.2 standard deviations of the logit of the propensity score (11). The balance between the 2 groups after matching was checked graphically and descriptively. A standardized difference of less than 10% indicates an appropriate balance (11). Mean standardized mean differences (SMD) with their range across the 30 multiple imputed datasets are presented in Figure 2 and Table 2 (see Supplementary data).
Figure 2. Love plot showing mean standardized mean differences (between TFNA and GTN3) with their range across the 30 multiple imputed datasets before (unadjusted) and after propensity-score matching (adjusted).
On each of the 30 imputed and propensity score-matched datasets, we used Cox regression to analyze the effect of intramedullary nail on failure of fixation (12). The resulting hazard ratios (HRs) and absolute risk differences were pooled using Rubin’s rule. Absolute risk differences were estimated at 3 years’ follow-up. 95% confidence intervals (CI) were calculated for the HRs and absolute risk differences.
As the TFNA was newly introduced in our institution, we used the exponential weighted moving average (EWMA) method to assess a potential learning curve. EWMA is a well-known statistical process monitoring tool because of its exceptional pace in catching infrequent variations in the process parameters, and is commonly used to monitor hospital data (13).
By the choice of weighting factor, λ, the EWMA procedure can be made sensitive to either detect a small change or a trend in the process. We used a λ of 0.005 to discern a trend in the observed outcomes (14). We performed a learning curve analysis on overall fixation failures, operation (OR) time, cut-out, and malpositioning. Besides overall fixation failures, OR time was chosen because a relatively prolonged operation time is anticipated on the introduction of a new implant. Both cut-out and malpositioning were chosen from the subcategories for a learning curve analysis as for these 2 in particular a potential learning curve effect can be anticipated. In addition, we performed sensitivity analyses excluding the first 50, 150, and 250 TFNA cases to assess the influence of a potential learning curve on the results.
Statistical analyses were performed using R 3.6.3 (R Foundation for Statistical Computing, Vienna, Austria) using packages “mice,” “cobalt,” “MatchIt,” and “riskRegression.”
Ethical approval for the study was granted by the Institutional Review Board (decision 2020-1596). The datasets generated during and/or analyzed during the current study are available from the corresponding author on reasonable request. No funding was obtained and the authors have no conflicts of interest to declare.
1,377 intramedullary nails were implanted in 1,353 consecutive patients. In 562 (556 patients) trochanteric fractures a TFNA was placed, and 815 (797 patients) trochanteric fractures were treated with a GTN3. After exclusion of 20 patients in the TFNA group and 18 patients in the GTN3 group due to a pathologic fracture, 542 (536 patients) TFNAs and 797 (779 patients) GTN3s were included in this study. Fixation failure rates for both groups are presented in Figure 3.
Figure 3. Cumulative incidence with 95% CI of failure of fixation for TFNA and GTN3. P-value = 0.002 (log-rank test). For TFNA and GTN3 abbreviations, see Table 1.
The Cox regression analysis revealed a failure of fixation of 14% in the TFNA group versus 7.0% in the GTN3 group, resulting in a hazard ratio (HR) of 2.0. The absolute risk difference of fixation failure in the TFNA group was 6.8% higher at 3 years’ follow-up compared with the GTN3 group (Table 3).
Factor | HR (95%CI) | TFNA (%) | GTN3 (%) | Risk difference (% (95%CI)) |
Overall failure | 2 (1.2-3.5) | 14 | 7.0 | 6.8 (1.7 to 12) |
Cut-out | 2.2 (0.9–5.7) | 4.8 | 2.2 | 2.6 (0.1 to 5.2) |
Implant breakage | 1.1 (0.1-9.0) | 0.6 | 0.6 | 0.0 (-1.2 to 1.2) |
Non-union | 0.9 (0.3-3.3) | 1.6 | 1.7 | -0.2 (-2.4 to 2.1) |
Malpositioning | 4.7 (0.7-34) | 1.5 | 0.3 | 1.2 (0.0 to 2.5) |
Fracture around nail | 4 (1.0-16) | 5.2 | 1.4 | 3.9 (-0.8 to 8.5) |
Miscellaneous | 1.2 (0.2–8.2) | 1.2 | 1.0 | 0.1 (–1.9 to 2.2) |
For TFNA and GTN3 abbreviations, see Table 1. |
The failures in the TFNA group versus the GTN3 group for each subgroup were: cut-out in 4.8% and 2.2%, implant breakage in 0.6% and 0.6%, non-union in 1.6% and 1.7%, malpositioning in 1.5% and 0.3%, new fracture around the nail in 5.2% and 1.4%, and miscellaneous in 1.2% and 1.0% (Table 3).
The TFNA EWMA plots revealed fluctuating curves around the mean without a clear decreasing trend in time for all 4 outcomes, indicative of the absence of a clear learning curve effect in the TFNA group (Figure 4). A separate sensitivity analysis excluding the first 50, 150, and 250 TFNA cases showed our results for fixation failure of the TFNA versus the GTN3 to be robust (Table 4, see Supplementary data).
Figure 4. Exponentially weighted moving average (EWMA) plots for overall failures, implant cutout, malpositioning, and operation time for both TFNA and GTN3. Dotted lines illustrate the mean failure rates during the study period. For TFNA and GTN3 abbreviations, see Table 1.
This study revealed an increased risk of fixation failure in the TFNA group. We found a hazard ratio (HR) of 2.0 (CI 1.2– 3.5) for fixation failure in the TFNA group compared with the GTN3 group. Cut-out and malpositioning appeared to be the most commonly encountered failure mechanisms. Learning curve analyses revealed that the potential influence of a learning curve could only have been minimal.
The TFNA was introduced to the global market with a number of innovative implant modifications to improve clinical performance. To date, literature regarding the TFNA is limited to a few small case series reporting fixation failure rates of 2–7% (15-18). Amongst these case series, 2 recent studies also reported concerns about mechanically failed TFNA implants (17,18).
The higher risk of fixation failure in the TFNA group we found was mainly attributed to a higher risk of cut-out, malpositioning, and new fracture around the nail. In particular, the failure mechanisms cut-out and malpositioning may have been influenced by surgical technique. As learning curve analyses revealed no clear learning effect the encountered increase in fixation failures appears to be at least partly implant related. Several different features were introduced at a time within the TFNA such as a different anatomical fit, a smaller proximal diameter (LATERAL RELIEF CUT design), an oblique cut on the lateral end of the blade/screw, and a different design of the proximal hole (BUMP CUT design) (7). Lastly, the blade has been introduced to replace the calcar screw. Biomechanical studies presented in the manufacturer’s value analysis brief reported a significantly higher resistance to cut-out of the TFNA helical blade compared with the TFNA screw (7,19).
2 clinical studies using the TFNA reported on higher rates of implant cut-out with the use of a helical blade compared with lag-screw fixation in the treatment of trochanteric fractures with intramedullary devices (20,21). These reports are in line with our study where 4.8% fixation failure was encountered in the TFNA group from implant cut-out versus 2.2% in the GTN3 group (Table 3). Of note is that a blade was used in 74% of our cases in the TFNA group because an age cut-off above 70 years old was used for the blade. As such, one may argue whether this concept of a blade versus a screw is indeed an improvement. In fact, the blade is hammered into the femoral head and so fracture distraction is introduced whereas compression is commonly accepted to increase chances of fracture healing. Furthermore, the oblique cut on the lateral end of the blade/screw was introduced to reduce lateral protrusion on the soft tissues (7). One could argue that the higher risk of malpositioning encountered in the TFNA group may partly have been the consequence of this innovation as the blade/screw can now only be inserted in the femoral head with steps of 360 degrees of rotation to align the lateral cortex while before more subtle 90-degree steps of insertion were an option.
Our study has limitations. Due to the retrospective design, confounding is likely to be present. In an attempt to minimize this potential confounding, we performed propensityscore matching of patients on important confounding factors. Despite matching of patients on a subset of baseline characteristics, differences in unmeasured covariates probably still exist. For example, both an inter- and intra-surgeon difference concerning indication for reoperation cannot be ignored. This potential bias appears to be limited, as indication for reoperation was established by consensus between all orthopedic/ trauma surgeons in weekly trauma meetings throughout the entire study period, thus avoiding single-surgeon decisions. Nevertheless, a randomized controlled trial is the ideal tool to control for bias. Furthermore, relatively small absolute differences are found in fixation failure rates between groups, and one may argue whether these differences are clinically relevant. Nonetheless, modifications in implant designs are aimed to improve clinical results and as such a decrease in clinical outcome, irrespective of a small margin, does not match with true gain in sustainability. Lastly, we did not correct for correlated bilateral cases in the analysis, while the methods of our statistical analysis do assume independent observations. However, previous studies showed bilateral surgeries do not introduce dependency problems in register studies (22,23).
From the limited data available in the literature on TFNA together with the encountered relatively high risk of fixation failure in this study there is reason to raise questions concerning the clinical benefit of the modifications applied to this device. In general, there is a tendency to overestimate the benefit of a newly introduced device (24). Newly introduced implants are often embraced by professionals after strong marketing without solid evidence (25). This rapid uptake of newly introduced implants into the market is, however, not without risks, in particular patients’ health-related risks. For this reason, a structured introduction and regulated clinical monitoring of implant innovations has repeatedly been recommended (25). Such regulation and monitoring may also apply to trauma devices for internal fixation. In hindsight, we regret that also in our institution a new implant device was introduced without solid clinical evidence behind it. An institution-driven larger change of (trauma) implant supplier was the driving force behind this. This study may help to underline the need for medical doctors with a critical and scientific background to be at least involved in implant choices.
In conclusion, this study revealed an increased risk of fixation failure in the TFNA proximal femoral nail in comparison with the Gamma nail (GTN3) for the treatment of trochanteric fractures. Cut-out and malpositioning were the most commonly encountered failure mechanisms. Modifications in implant design may have played a role in this increased risk of failure of fixation, because no clear learning curve effect was found.
Balance across imputations before PSM | Balance across imputations after PSM | |||||
Characteristic | TFNA | GTN3 | Mean SMD (range) | TFNA | GTN3 | Mean SMD (range) |
Mean age (SD) | 78.8 (13.5) | 79.9 (11.8) | –0.09 (–0.09 to –0.09) | 79.0 (12.7) | 79.2 (12.7) | –0.02 (–0.06 to 0.02) |
Female, % | 70 | 74 | –0.09 (–0.09 to–0.09) | 71 | 72 | –0.04 (–0.10 to –0.01) |
ASA score, % | ||||||
1 | 5.9 | 4.5 | 0.06 (0.05 to 0.07) | 5.6 | 5.8 | –0.003 (–0.06 to 0.05) |
2 | 25 | 41 | –0.36 (–0.37 to –0.35) | 28 | 30 | –0.04 (–0.10 to –0.01) |
3 | 51 | 45 | 0.068 (0.06 to 0.08) | 56 | 54 | 0.04 (–0.02 to 0.10) |
4 | 17 | 4.6 | 0.39 (0.37 to 0.40) | 8 | 7.8 | 0.01 (–0.04 to 0.03) |
5 | 1.8 | 1.6 | 0.01 (–0.01 to 0.02) | 2.1 | 2 | 0.010 (–0.03 to 0.06) |
Mean BMI (SD) | 24.1 (4.4) | 24.0 (4.2) | 0.04 (0.02 to 0.08) | 24.1 (4.4) | 24.1 (4.3) | 0.01 (–0.05 to 0.05) |
Smoking, yes, % | 17 | 26 | –0.23 (–0.29 to –0.18) | 18 | 18 | –0.01 (–0.07 to 0.03) |
Fracture type, % | ||||||
A1 | 38 | 36 | 0.03 (0.03 to 0.03) | 40 | 38 | 0.04 (–0.01 to 0.01) |
A2 | 43 | 41 | 0.05 (0.05 to 0.05) | 40 | 41 | –0.02 (–0.08 to 0.02) |
A3 | 15 | 16 | –0.05 (–0.05 to –0.05) | 15 | 16 | –0.02 (–0.07 to 0.05) |
Subtrochanteric | 4.1 | 6 | –0.10 (–0.10 to –0.10) | 4.6 | 4.7 | –0.01 (–0.08 to 0.03) |
Size, % | ||||||
Short | 77 | 78 | 0.023 (0.02 to 0.02) | 78 | 78 | –0.001 (–0.07 to 0.05) |
Long | 23 | 22 | 22 | 22 | ||
Balance between groups is presented as mean standardized mean differences (SMD) with the corresponding minimum and maximum standardized mean differences across the 30 multiple imputed datasets. Patient characteristics are presented as pooled mean (SD) for continuous variables or as pooled percentages for binary or categorical variables across multiple imputations. For TFNA and GTN3 abbreviations, see Table 1. |